6
Installation test
18
ÄKTAexplorer Installation Guide 18-1139-59 Edition AC
6 Installation test
The installation test is designed to check the function of the liquid
delivery and the UV monitoring system of ÄKTAexplorer. The
installation test can also be used at any time to check the condition of
the system, e.g. after a prolonged stop.
Correct gradient formation is tested by producing a linear gradient and
a series of concentration steps of acetone.
Correct UV monitoring is tested by monitoring the acetone
concentration at 265, 254 and 280 nm and calculating the absorbance
ratios 265 nm/254 nm and 265 nm/280 nm.
6.1
Preparation of ÄKTAexplorer
6.1.1
Startup of the ÄKTAexplorer separation unit
1
Switch on the separation unit using the mains switch located to the
left on the base platform.
6.1.2
Startup of the computer and UNICORN software
1
Switch on the computer, the display and the printer according to the
instructions in the manufacturer manuals.
2
Log into Windows
®
by first pressing
Ctrl-Alt-Del
, and then clicking
OK
.
3
When the Windows desktop appears, start UNICORN by double-
clicking on the UNICORN icon.
4
Select user
default
and enter
default
as password. Click
OK
.
ÄKTA
explorer
Mains switch
18113959AC_TK.book Page 18 Monday, June 23, 2003 4:45 PM