89
98-0157 Rev K
DUAL PASS:
‘Default’ refers to the settings that will be entered if the “Install Factory Defaults”
button is pressed
‘Initial Setup’ refers to the settings that are entered in the system when it leaves
the factory.
‘User Settings’ refer to the settings on the unit when it is put into service. This
should be filled out when the unit is put into service and should be updated any
time settings change
Setting
Default Initial Setup User Setting
Units
Setting
Default Initial Setup User Setting
Units
Auto Flush
1
min
R01 Concentrate offset
217.3
Auto Rinse Min
1
min
R01 Concentrate Gain
1.132
Auto Rinse Max
10
min
R02 Concentrate Offset
217
Target Recovery Stage 1
65
%
R02 Concentrate Gain
3.487
Target Recovery Stage 2
85
%
R02 Permeate Offset
316
Standby Refresh Time
10
min
R02 Permeate Gain
5.6
Standby Refresh Interval
120
min
Interstage Pressure Offset
0
Interstage Pressure Gain
1
Setting
Default Initial Setup User Setting
Units
RO1 Concnetrate Valve Limits Min
15
Tank Temp
185
°F
RO1 Concnetrate Valve Limits Max
75
Tank Return Temp
180
°F
RO2 Concnetrate Valve Limits Min
15
Hold period
240
min
RO2 Concnetrate Valve Limits Max
75
Setting
Default Initial Setup User Setting
Units
Setting
Default Initial Setup User Setting
Units
Clean Reminder
1
days
Conc Control Valve 1 RO Heat
Disinfection
0
Recirc Period
25
min
Conc Control Valve 2 RO Heat
Disinfection
0
Rinse Pressure Period Low
40
min
Conc Control Valve 1 Chem Clean
0
Rinse Pressure Period High
15
min
Conc Control Valve 2 Chem Clean
0
Setting
Default Initial Setup User Setting
Units
Setting
Default Initial Setup User Setting
Units
Tank Offset
0
Loop Over temp Warning
86
°F
Tank Gain
90
Loop Over temp Alarm
95
°F
Stage 1 Offset
5
RO Heat Disinfect Overtemp Tank
194
°F
Stage 1 Gain
1.5
RO Heat Disinfection Over Temp
Permeate 2
183
°F
Output Offset
0
RO Heat Disinfection Over Temp
Tank return
183
°F
Output Gain
0.95
R02 Pressure setpoint Low
14
psi
RO2 Pressure setpoint High
725
psi
Setting
Default Initial Setup User Setting
Units
Loop Supply Flow Setpoint Low
1
gpm
RO1 Target Recovery
75
%
Loop Supply Flow Setpoint High
18
gpm
RO2 Target Recovery
85
%
Low Pressure Alarm Delay
0
hrs
Dead Band
20
%
RO1 Flush Valve Position
75
%
Setting
Default Initial Setup User Setting
Units
Cool Down End Temp
86
°F
Setting
Default Initial Setup User Setting
Units
Pump Run Slow Temp
90
°F
Tank Quality Warning
1100
uS
Tank Quality Alarm
1200
uS
Setting
Default Initial Setup User Setting
Units
Output Quality Warning
30
uS
Chem Clean Rinse
45
min
Output Quality Alarm
100
uS
RO Clean
Line Cell Calibration
RO Parameters
Alarm Setup
Engineers Alarm Setup
Engineers RO Heat Disinfection Setup
Engineers Chem Clean Setup
MediQA Settings DPH
RO Supply
Flow Regulation
Ro Heat Disinfection
Concentration Valve Settings
Содержание MediQA MDP Series
Страница 70: ...69 98 0157 Rev K 10 FLOW DIAGRAMS SPH UNITS ...
Страница 71: ...70 98 0157 Rev K DPH UNITS ...
Страница 72: ...71 98 0157 Rev K 11 WIRING SCHEMATIC SPH UNITS ...
Страница 73: ...72 98 0157 Rev K ...
Страница 74: ...73 98 0157 Rev K ...
Страница 75: ...74 98 0157 Rev K ...
Страница 76: ...75 98 0157 Rev K ...
Страница 77: ...76 98 0157 Rev K ...
Страница 78: ...77 98 0157 Rev K ...
Страница 79: ...78 98 0157 Rev K ...
Страница 80: ...79 98 0157 Rev K DPH Units ...
Страница 81: ...80 98 0157 Rev K ...
Страница 82: ...81 98 0157 Rev K ...
Страница 83: ...82 98 0157 Rev K ...
Страница 84: ...83 98 0157 Rev K ...
Страница 85: ...84 98 0157 Rev K ...
Страница 86: ...85 98 0157 Rev K ...
Страница 87: ...86 98 0157 Rev K ...