Calibration of Sensors
Login at USER 5 level.
The password will only be available to
authorized personnel; contact AmeriWater if
this level of access is required.
Enter test screen
FACTORY SETUP PAGE 6
and select
“
CALIBRATE
”.
Three sensor options are displayed.
Identify the sensor to be calibrated.
Adjust the “Offset “
To check/adjust calibration of the
temperature sensors the actual water
temperature should be measured using an
independent calibrated device.
Adjust the Gain and Bias such that the
sensors match the reading of the
independent device.