AmeriWater DISTRIBUTION PUMP SYSTEM Скачать руководство пользователя страница 6

98-0096A 

4

6.0  ADJUSTING THE TIMER 

With the power off and the system properly locked out, open the control panel.  In the middle 
of the box, you will see the OMRON timer as shown below.  To adjust the timer, simply insert 
a small screwdriver into the screw on the lower right hand face of the timer.  Rotating this will 
toggle between the time interval modes, which are displayed on the face of the timer.    

Time Interval 

Adjustment 
Screw 

Once all validation of your set-up is complete, you must re-set the timer to 5 hours to meet the 
AAMI standards. 

Содержание DISTRIBUTION PUMP SYSTEM

Страница 1: ...SSV DISTRIBUTION PUMP SYSTEM OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414 P N 98 0096 REV A...

Страница 2: ...0 INTRODUCTION 1 2 0 THEORY OF OPERATION 1 3 0 INSTALLATION 2 4 0 START UP 2 5 0 DAILY MONITORING 3 6 0 ADJUSTING THE TIMER 4 7 0 SPARE PARTS LIST 5 8 0 PUMP CURVES 6 9 0 SYSTEM DRAWINGS 8 10 0 TROUB...

Страница 3: ...is process Running the distribution pump for an extended period without water will result with overheating and damage to the pump The distribution pumps are available as a single pump or a dual pump A...

Страница 4: ...on pump wiring to the motor contactor located in the control panel refer to the control wiring diagram provided in this manual and with the distribution pump control Connect the three phase or single...

Страница 5: ...that the low level cut off switch operates correctly For a dual pump system this will need to be done for each pump 8 For the dual pump system only the timer will need to be verified This is accompli...

Страница 6: ...shown below To adjust the timer simply insert a small screwdriver into the screw on the lower right hand face of the timer Rotating this will toggle between the time interval modes which are displaye...

Страница 7: ...3 Pump 3HP Three Phase 60GPM 80 SV04 Check Valve Ball Spring Loaded 041 0033 Stainless Steel Pressure Gauge 0 100psi 43530710 Contactor AB 9 AMP 3 Pole 120V Coil 61670245 Motor Contactor Overload 12 A...

Страница 8: ...98 0096A 6 10 0 PUMP CURVES...

Страница 9: ...98 0096A 7...

Страница 10: ...98 0096A 8 9 0 SYSTEM DRAWINGS Single Distribution Single Phase Controller...

Страница 11: ...98 0096A 9 Single Distribution Single Phase Controller Schematic...

Страница 12: ...98 0096A 10 Single Distribution Three Phase Controller...

Страница 13: ...98 0096A 11 Single Distribution Three Phase Controller Schematic...

Страница 14: ...98 0096A 12 Dual Distribution Controller...

Страница 15: ...98 0096A 13 Dual Distribution 220 Volt Single Phase Controller Schematic Dual Distribution 208 Volt Three Phase Controller Schematic...

Страница 16: ...98 0096A 14...

Страница 17: ...98 0096A 15 Dual Distribution 230 460 Volt 3 Phase Controller Schematic...

Страница 18: ...98 0096A 16 Single Distribution Pump Assembly...

Страница 19: ...98 0096A 17 Dual Distribution Pump Assembly...

Страница 20: ...e at each station one valve at a time until water begins to escape When water begins to escape close the valve and proceed to the next station This will purge all of the air from the loop Verify that...

Страница 21: ...rnia to cause cancer birth defects or other reproductive harm One such chemical is Vinyl Chloride a compound used to produce Polyvinyl Chloride PVC The AmeriWater system you have purchased may contain...

Отзывы: