Appendix C:
Derating Tables
Before setting up the GPS350, or GPS360 interrupter, consider the outside temperatures (both high
and low) that the interrupter will be exposed to for the duration of your survey. For each
temperature maximum, American Innovations recommends a maximum rectifier current for a given
interruption cycle.
The following tables identify the maximum rectifier current that is recommended when using the
interrupter in the corresponding maximum outside air (ambient) temperature. Refer to these tables
when setting up the GPS350, or GPS360 interrupter.
l
GPS350 Derating Table on page 59
l
GPS360 Derating Table on page 59
GPS350 Derating Table
Before setting up the GPS350 interrupter, consider the outside temperatures (both high and low)
that the interrupter will be exposed to for the duration of your survey. For each temperature
maximum, American Innovations recommends a maximum rectifier current for the given
interruption cycle.
Ambient Temperature
Interruption Cycle
Celsius
Fahrenheit
0.8 s ON — 0.2 s OFF
4 s ON — 1 s OFF
25°
77°
60 A
80 A
35°
95°
60 A
80 A
45°
113°
58 A
78 A
55°
131°
49 A
66 A
65°
149°
38A
51 A
Table C-1. GPS350 Derating
GPS360 Derating Table
Before setting up the GPS360 interrupter, consider the outside temperatures (both high and low)
that the interrupter will be exposed to for the duration of your survey. For each temperature
maximum, American Innovations recommends a maximum rectifier current.