32
-
2. Draining the Chamber
●
When pressure is remaining in the chamber, do not remove the drain bottle or drain the chamber,
otherwise hot water and steam may gush out, causing a scald.
●
When the unit is not to be used for a long time, be sure to drain the chamber to prevent pipe
clogging. Confirm that the inside of the chamber has cooled sufficiently before draining the
chamber.
①
Open the drain valve at the bottom of the left
side of the unit, and the water in the chamber will
be poured into the drain bottle.
②
When the drain bottle is full of water, the drain
bottle display blinks in red, and an electronic
alarm sounds (peep). Close the drain valve, and
drain the bottle.
③
Repeat the operation several times until the
water inside the chamber disappears.
・
Close the drain valve completely.
3. Cleaning inside the Chamber
●
The heater at the bottom of the chamber is provided with a temperature sensor.
Exercise care so that you will not have your finger cut by the edge during cleaning.
①
Take out the bottom plate, and check the chamber bottom and heater surface for dirt.
Clean them with a soft brush and the like after draining.
②
When the temperature sensor is off the clamping clips, attach the temperature sensor
correctly to the clamping clips.
HG-50(with 2kw heater) HG-50 with 3kw heater & HG-80
CAUTION
Bottom at left side
Closed
Open
Drain valve
CAUTION
Heater
Temperature sensor
Clamping clips
Heater
Temperature sensor
Clamping clip