background image

 

AMD General Examination Camera 

AMD Telemedicine 

Version 1.0 

Zoom Lens 

The AMD General Examination Camera has two interchangeable lenses: 

 

 

 

The Protective Lens provides protection for the 
camera lens and the ring-light. 

 

The 50x Zoom Lens provides 50x magnification for a 
closer look at objects. This lens has a flexible rod that 
indicates the focal length of the lens. When you use 
this lens, place the tip of the rod gently on the object 
to be viewed, and the camera will be at the correct 
viewing distance. 

The AMD-2505 Disposable Tip Covers can be used to 
protect the flexible tip of the 50x lens against 
contamination.  These tip covers should be replaced/  
disposed of properly as local rules govern after each 
patient contact.  These tip covers “ roll out”  condom-
style for installation. 

* Local rules may require these contaminated tip 
covers to be treated as hazardous waste. 

 

To remove a lens: 

 

1.

 

Place your fingers in the indent on the lower left corner 
of the lens. 

2.

 

Pop the lens off. 

 

To attach a lens: 

                

 

1.

 

Position the lens on the front of the camera. 

2.    Snap the lens in place. 

 

Содержание AMD-2500

Страница 1: ...AMD Telemedicine Version 1 0 AMD 2500 General Examination Camera AMD Telemedicine 67 Middle Street Lowell MA 01852 800 742 1674 978 937 9021 www amdtelemedicine com...

Страница 2: ...AMD General Examination Camera Using the Camera Setting Up the Camera Displaying an Object Adjusting the Image Sorting Out the Details Connectors and Controls Cable Connections Zoom Lens AMD 2500 Prod...

Страница 3: ...can be used for many applications including dermatology and dentistry The small size and built in controls allow for unobtrusive examination of various areas of the body Image Control Buttons Protecti...

Страница 4: ......

Страница 5: ...a make sure that the DC VBS cable is secure 3 Plug the following three cables into the AMD junction box a DC VBScable b RCA video cable c Power adapter cord 4 Plug the RCA connector of the video cable...

Страница 6: ...part WA112M 6 Move the Power switch to ON move the Light switch to ON and verify that the green power indicator light is ON 7 The AMD 2500 should be inspected for proper operation daily The AMD 2500 s...

Страница 7: ...the image of the white sheet c Press the WB button on the back of the camera for five seconds The colors are automatically rebalanced For more information about White Balance refer to Sorting Out the...

Страница 8: ...l Examination Camera AMD Telemedicine Version 1 0 Notes 5 The AMD 2500 should be inspected for proper operation daily The AMD 2500 should be cleaned disinfected after each patient contact at a minimum...

Страница 9: ...ch definition in the image 1 Turn ON the Ring Light using the switch on the back of the Examination Camera 2 Press the REF button on the top of the Examination Camera to toggle the Reflection mode ON...

Страница 10: ...AMD General Examination Camera AMD Telemedicine Version 1 0 Notes If the image is too large 1 Zoom out by pressing the W button on the top of the Examination Camera...

Страница 11: ...the FRE button again If the image is too dark 1 Turn ON the Ring Light using the switch on the back of the Examination Camera 2 If the image is still too dark make sure that the Examination Camera is...

Страница 12: ...till too pale make sure that the Examination Camera is using Automatic Iris Control The iris indicator light on the back of the Examination Camera should be OFF 3 If the image is still too pale a Pres...

Страница 13: ...t the end of the flexible rod 2 Make sure that the Examination Camera is using Automatic Focus Control The Focus switch should be set to AF 3 If the image is still fuzzy a Move the Focus switch to MF...

Страница 14: ......

Страница 15: ...2505 Disposable Tip Cover to prevent contamination Tele Button zooms in on objects For the procedure for using this button refer to the If the image is too small section Reflection Button toggles the...

Страница 16: ...AMD General Examination Camera AMD Telemedicine Version 1 0...

Страница 17: ...set to Manual Focus the FARand NEAR buttons are used to focus on objects F FARmanually focuses the camera on distant objects G NEARmanually focuses the camera on close objects H Manual Iris Indicator...

Страница 18: ...There are two connections from the junction box the RCA cable plugs into the External Video Input of the video conference system and the power cord plugs into an outlet For details on hooking up the...

Страница 19: ...t to be viewed and the camera will be at the correct viewing distance The AMD 2505 Disposable Tip Covers can be used to protect the flexible tip of the 50x lens against contamination These tip covers...

Страница 20: ...AMD General Examination Camera AMD Telemedicine Version 1 0...

Страница 21: ...t ships with the product to ensure labelled safety standards are met This power supply is manufactured by Jerome Industries part WA112M Camera Dimensions w x d x h 2 2 x 3 7 x 2 9 5 588cm x 9 398cm x...

Страница 22: ...AMD General Examination Camera AMD Telemedicine Version 1 0 Power Supply Safety Ratings Type BF equipment Class II equipment Read all instructions provided with product Indoor use only...

Страница 23: ...AMD General Examination Camera AMD Telemedicine Version 1 0 AMD 2500 System Configuration Diagram AMD 2500 DC VBS Cable Junction Box RCA Cable AC Adapter Cable TV Monitor Ferrite Ferrite...

Отзывы: