background image

6.

7.

Insertion of the aScope 3 & aScope 3 Slim

Removal procedure

  Slowly withdraw the aScope while observing the image on the monitor.  

The distal tip must be in a neutral and non-deflected position.

  After use disconnect the aScope from the aView and dispose of it in accordance with 

local guidelines for collection of infected medical devices with electronic components.

  Wipe down the aView as per the instructions in the Instructions For Use.

  The tube connection can be used to fixate 

the selected ET tube during intubation.

  Keep the cord of the aScope 3 or  

aScope 3 Slim as straight as possible by 
holding the handle of the aScope 3 high. 
This increases the range of movement of 
the tip and allows you more control.

  Advance the aScope 3 or aScope 3 Slim 

slowly, allowing the light intensity of the 
image to re-adjust.

  If the image is affected by blood or 

secretion you can try to clear the 
secretion by flicking the bending lever up 

and down or dabbing the camera tip 
gently against the mucosal wall to clear 
the image.

  If the above fails, remove the aScope and 

rub the lens clean with either gauze or an 
alcohol swap. 
Do not be afraid to exert a little pressure 
when doing this. Proceed as above.

Ambu A/S

Baltorpbakken 13 • DK-2750 Ballerup • Denmark
T +45 72 25 20 00 • F +45 72 25 20 50
www.ambu.com

4

96 3

60

3 0

1 - 

V0

1 - 2

0

13

/0

9 - 

Ambu 

A/S. 

Tec

hnic

al da

ta ma

y b

e mo

dified w

ithou

t f

ur

ther no

tic

e.

  The aView can be placed on a flat surface or secured to an I.V. pole using the  

bracket supplied. The bracket also has a hook that facilitates easy, accessible  
storage of the aScope 3 via the hole in the top of the aScope 3/aScope 3 Slim pouch.

  Most users find the default contrast and brightness settings to be optimal but  

these can be adjusted if required.

  Consider dimming the room lights to see the image from the aScope 3 or  

aScope 3 Slim  more clearly.

  Check the battery indicator on the aView. 

If the time remaining is not sufficient for the procedure charge the monitor by 
plugging it into the mains.

4.

5.

Preparation for use

Icons

Max. battery status 
of the aView

Min. battery status 
of the aView

The icon remains white until one block is 
left, after which it turns red.
When remaining battery capacity is 10% 
the red battery icon starts flashing.

Charging is shown with blocks flashing.
Current capacity is shown with non-
flashing blocks.

Battery is charging

Battery current 
capacity

If the battery is fully charged and still 
connected to a charger the battery 
icon changes to:

Start-up picture

Real time image will 
appear after only a  
few seconds.
After appr. 45 seconds 
the user Interface is 
ready displaying the 
icons for battery status, 
brightness/contrast 
control, device usage 
time  as well as the as 
snapshots and  
recording of videos.

When the aView is 
turned on the ON/OFF 
button will light up 
green. During charging 
the ON/OFF button will 
light orange.

Отзывы: