background image

Dr A Ruthven – Version 1.2 – November 2021 

Page 2 of 2 

 

PROGRAMMING THE PUMP

 

1.  Open the white battery cap by twisting anticlockwise and pulling. Insert 2 x new AA batteries. 
2.  Turn on the pump – replace the white battery cap and twist it clockwise until black and white lines align. 
3.  The pump will be in pause mode – a flashing 

;

 will be displayed. 

4.  Clear the pump history (READ THIS ENTIRE STEP BEFORE STARTING): 

  Press and hold 

  

for 2 seconds – pump beeps & displays ‘Hx’. 

  Release and immediately then re-press and hold   for 2 seconds – pump beeps & displays ‘Hx O’.  

  The history has now been cleared. If ‘Hx O’ was not displayed, repeat this step until it is. 

5.  The pump remains in pause mode – a flashing 

;

 will be displayed. 

6.  Press and hold 

both

 

 

and 

 

 together for 2 seconds until a beep is heard & PRO is displayed briefly. 

7.  Flashing PIb should be displayed. If not, press 

 

to cycle through modes until PIb is displayed. 

8.  Press and hold 

  

until a beep is heard to select PIb mode. 

9.  Set the desired dose (d) i.e. programmed intermittent bolus volume by 

tapping

 

 

to increase /   

 

to 

decrease (see prescription charts for suggested doses). 

Do not press and hold buttons to change faster.

 

10.  Press and hold 

  

until a beep is heard to confirm the dose & repeat for the decimal. 

11.  Int will be displayed then disappear. Set the bolus interval to 4:00 (4 hours) in the same way (tap 

 

and 

  

to change, press and hold 

  

to confirm). 

12.  Set the ‘volume to be infused’ to the same as the volume of the attached local anaesthetic bag. 

Note volumes vary between drug manufacturers – please check carefully when programming pump. 

13.  A flashing 

;

 

signifies programming is complete. The pump remains in pause mode. 

14.  Press 

 

 

to review programming. The settings will be displayed then the pump resumes pause mode. 

15.  If you have made an error at any point you must repeat the entire process (go back to step 6). 

 

STARTING THE PROGRAM 
Do this ASAP after loading dose given (even if pump not yet attached to catheter) 

1.  Press 

 to start the program. If you have not already reviewed the program settings, these will be 

displayed in sequence and you then need to press 

 

again to start the program. 

2.  A flashing green light under 

  

shows that the program is running. The display should show 0.0ml. 

Do not press 

  

within 60 seconds of starting the program or an initial bolus will be delivered. 

If 

  

is pressed in error and an inadvertent bolus is started, see box below.

 

3.  If the display shows a volume other than 0.0ml, the history has not been successfully cleared. Press 

 

to pause the pump and repeat step 4 under ‘programming the pump’ above, then press 

 to restart. 

4.  Place pump & local anaesthetic in black carry bag, being careful not to kink tubing within the bag. 
5.  Document date/time program started on prescription chart. 

 

REVIEWING THE CURRENT PROGRAM SETTINGS 

1.  Press 

 to pause the pump – a flashing 

;

 

is displayed. 

2.  Press 

 

 – the display shows ‘Rx’ then cycles through: 

  Program type (should always be set to PIb) 

  Programmed intermittent bolus volume (e.g. 20-30ml) 

  Dosing interval (should always be set to 4:00 i.e. 4 hours) 

  Volume to be infused remaining 

3.  The pump remains in pause mode with a flashing 

;

 

displayed. 

4.  Press 

 

 

to restart the program. 

  

flashes green to show the pump is running & the display shows 

the total volume infused since the history was last cleared. 

 

 

STOPPING

 

AN

 

INADVERTENT

 

INITIAL

 

BOLUS

 

1.  Press 

  to enter pause mode and stop bolus administration. A flashing 

;

 

is displayed. 

2.  Clear the pump history (READ THIS ENTIRE STEP BEFORE STARTING): 

  Press and hold 

  

for 2 seconds – pump beeps & displays ‘Hx’. 

  Release and immediately then re-press and hold   for 2 seconds – pump beeps & displays ‘Hx O’.  

  The history has now been cleared. If ‘Hx O’ was not displayed, repeat this step until it is. 

3.  Press 

 

 

to restart program. 

  

flashes green to show the pump is running & the display shows 0.0ml. 

Отзывы: