CALIBRATION OF BLANK HOLDER D-SET

E N

2

Calibration of blank holder 
D-Set

Before using the workpiece holder D-Set (181365) 
in the hybrid milling machine, it must be calibrated 
first

2.1

Product description

Calibratable D-Set workpiece holder for use in 
hybrid milling machines. The blank holder is used 
for fixation of prefabricated D-set blanks.

The blank holder consists of two components:

_

Blank holder

_

Fixation plate

2.2

Delivery scope

Fig. 1

Workpiece holder D-Set Motion 3

1

Blank holder (uncalibrated)

2

Fixation plate

3

Allen screws M4

×

10 (7 pce.)

The blank holder is delivered completely assem-
bled.

2.3

Required system components

_

CAD/CAM software for removable prosthetics 
(D-Flow)

_

Hybrid milling machine Ceramill Motion 3

_

Roto RFID 1.2 ST (751001)

In order to ensure the software functionality:

Before using, Install the current update of all 
software components.

NOTE: 

The workpiece holder is individually calibrated to 
the respective hybrid milling machine. After-
wards, the holder can no longer be used in other 
hybrid milling machines.

1

2

3

Содержание 181365

Страница 1: ...Gebrauchsanweisung 3 6 Instruction Manual 7 10 Kalibrierung Werkst ckhalter D Set 181365 Calibration of workpiece holder D Set 181365...

Страница 2: ......

Страница 3: ...nn eines Warnhinweises kennzeichnen Art und Schwere der Folgen falls die Ma nahmen zur Abwendung der Gefahr nicht befolgt werden _ HINWEIS bedeutet dass Sachsch den auftre ten k nnen _ VORSICHT bedeut...

Страница 4: ...rumfang Bild 1 Werkst ckhalter D Set Motion 3 1 Werkst ckhalter unkalibriert 2 Fixierblech 3 Innensechskantschrauben M4 10 7 St ck Der Werkst ckhalter wird fertig montiert gelie fert 2 3 Erforderliche...

Страница 5: ...erung des Werkst ckhalters durchf hren Bild 2 Werkst ckhalter D Set Motion 3 Werkst ckhalter ohne Fixierblech und Schrau ben in den Materialwechsler einlegen Werkzeug Roto RFID 1 2 ST 751001 in das We...

Страница 6: ...agung starten ausgew hlt wurde Job manuell an der Hybridfr smaschine star ten Es werden 16 Kalibriermarkierungen zur lage richtigen Fixierung der Zahnrohlinge gefr st Der Kalibriervorgang dauert ca 35...

Страница 7: ...substituted by a lightning bolt Signal words at the beginning of a warning indica tion specify the type and severity of the conse quences if the measures to avert the hazard are not adhered to _ NOTE...

Страница 8: ...2 Delivery scope Fig 1 Workpiece holder D Set Motion 3 1 Blank holder uncalibrated 2 Fixation plate 3 Allen screws M4 10 7 pce The blank holder is delivered completely assem bled 2 3 Required system c...

Страница 9: ...ig 2 Workpiece holder D Set Motion 3 Insert blank holder without fixation plate and screws into the material changer Insert tool Roto RFID 1 2 ST 751001 into the tool magazine The hybrid milling machi...

Страница 10: ...ansfer was not selected Start job manually at the hybrid milling machine 16 calibration marks are milled for fixation of the tooth blanks in the correct position The calibration process takes approx 3...

Страница 11: ......

Страница 12: ...n ISO 9001 Manufacturer Hersteller Distribution Vertrieb Amann Girrbach AG Herrschaftswiesen 1 6842 Koblach Austria Fon 43 5523 62333 105 Fax 43 5523 62333 5119 austria amanngirrbach com www amanngirr...

Отзывы: