or would incorrectly be displayed as 28640 if the totaliser had no decimal points. This could be
adjusted by changing the
ESCL
value to 2 i.e.
T otal
=
P revious total
+
3580
×
4
3600
×
10
2
×
7200
T otal
=
P revious total
+ 286(
m
3
)
If the total has more decimal point places than the rate then the adjustment would have to be the
other way i.e. multiplication by 10 or 100 etc. to adjust the result.
5.45
Negative total select
Display:
totI NEG
Range:
OFF
or
on
Default Value:
OFF
Displays and sets whether negative totals are allowed or not. When set to OFF negative totals
are not allowed and the total will not increase when the rate input is negative. Set to on to allow
negative totals.
5.46
Wrap around operation
Display:
totI RAP.F
Range:
ZERO
or
StOP
Default Value:
ZERO
Displays and sets the totaliser wrap around operation for displays at full scale. If
StOP
is selected
the display will halt at its maximum or minimum display value. If
ZERO
is selected then the
display will wrap around to zero i.e. will reset itself and start again at zero.
5.47
Clear total
Display:
CLR totI
Range:
n/a
Default Value:
n/a
Allows the totaliser value to be cleared via the setup pushbuttons located at the rear of the
instrument. To clear the totaliser press
^
and
v
simultaneously at this function. The message
CLrd
will be seen to confirm that the totaliser memory has been cleared. Alternatively the
P
button or a remote input can be used to clear the total. If a remote switch is to be used the
R.INP
function must be set to
CLR.t
. If the
P
button is to be used to clear the total the
P.but
function
must be set to
CLR.t
.
34 of
PM4IVTMAN-2.3-0