10
11
EMC guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The Alvita Infrared Ear Thermometer is intended for use in the electromagnetic environment specified below.
The user of the Alvita Infrared Ear Thermometer should ensure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The Alvita Infrared Ear Thermometer uses RF energy only
for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
The Alvita Infrared Ear Thermometer is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEX 61000-3-3
Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
The Alvita Infrared Ear Thermometer is intended for use in the electromagnetic environment specified below. The
customer or the user of the Alvita Infrared Ear Thermometer should ensure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
level
Electromagnetic environment-guidance
Conducted RF IEC
61000-4-6
3 Vrms
150KHz to
80Mhz
Not applicable
Portable and mobile RF communication
equipment should be used no closer to
any part of the RA600 series, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2 √P
d = 1.2 √P 80MHz to 800MHz
d = 2.3 √P 800MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, (a) should be less then the
compliance level n each frequency range.
(b) interference may occur in the vicinity
of equipment marked with the following
symbol:
Radiated RF IEC
61000-4-3
3V/m
80mHz to
2.5 GHz
3 V/m
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structure, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed FM transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Alvita Infrared Ear Thermometer is
used exceeds the applicable RF compliance level above, the Alvita Infrared Ear Thermometer should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Alvita Infrared Ear Thermometer.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and manufacturer’s declaration-electromagnetic immunity
The Alvita Infrared Ear Thermometer is intended for use in the electromagnetic environment specified below.
The user of the Alvita Infrared Ear Thermometer should ensure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
level
Electromagnetic environment-guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 9 kV air
± 6 kV contact
± 9 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
± 2kV for
power supply
lines
± 1kV for input/
output lines
Not applicable
Not applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5
± 1kV line(s) to
line(s)
± 2kV line(s) to
earth
Not applicable
Not applicable
Mains power quality should be that
of a typical commercial or hospital
environment
Voltage Dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5% UT (>95%
dip in UT) for
0,5 cycle
40% UT (60%
dip in UT) for 5
cycles
70% UT (30%
dip in UT) for 25
cycles
<5% UT (>95%
dip in UT for 5
seconds
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the Alvita
Infrared Ear Thermometer requires
continued operations during power mains
interruptions, it is recommended that
the Alvita Infrared Ear Thermometer be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
3 A/m
3 A/m
The Alvita Infrared Ear Thermometer
power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level
UK
UK