16 | 18
Appendix
Datasheet
Number of measuring channels:
1 (Strain gauge full bridge > 300 Ω)
Supply voltage:
12E - 10 … 18 VDC
24E - 18 … 30 VDC
Electronics protected against voltage reversal
Isolating proof voltage input to output:
200 V
Higher isolated proof voltage on request
Power consumption:
max. 3 W
Strain gauge excitation supply:
±2,5 VDC ± 1% / ±5 VDC ± 1%
Analogue output
0 … 10 V / ±10 V
4 … 20 mA
max. 1 mA (short-period short-circuit proof)
max. 500 Ω
Limit frequency (-3 dB):
1 kHz (10 Hz with low pass filter)
optional up to 30 kHz
Input resistance:
>3 MΩ
Max. input sensitivity:
100 mV/V at ±5 VDC excitation supply
Non-linearity:
±0.05 % FSO
Electrical connection:
Pluggable Screw clamps
Housing:
Plastic enclosure for top hat rail mounting (IP20)
Dimension (W x H x D):
23 x 99 x 115 mm
Weight:
approx. 150 g
Temperature, storage:
-20 °C … +60 °C
Temperature, operating:
0 °C … +50 °C
Order designation
SG-KP…
Single channel strain gauge-amplifier in plastic housing for DIN-top hat rail mounting (IP20)
…-12E-…
Supply voltage: 10 … 18 VDC
…-24E-…
Supply voltage: 18 … 30 VDC
…-010-…
Analogue output: 0 … 10 V
…-B10-…
Analogue output: ± 10 V
…-420-…
Analogue output: 0 … 10 V and 4 … 20 mA
… -GFxxx
frequency limit up to 30 kHz
…no declaration
standard version 1 kHz