6.3 Instrument Calibration
The system delivery comprises a calibration tile and test tiles for color and gloss,
see
Test and Calibration Standards [
. These tiles can be used to check if your
instrument is measuring correctly. Each tile comes with a dedicated serial number
matching your instrument. In the configuration menu select
Calibration
.
Information
Check date of last calibration, serial
numbers and values for calibration and
test tiles here.
White standard
Comprises the options for instrument
calibration.
Autodiagnosis
Comprises the options for auto diagno-
sis.
Calibrate on white standard
Start the manual calibration procedure
here.
Docking station interval
Select interval for the auto diagnosis
here. Possible values: 1, 3, 7, 14 days.
Default: 7 days.
Add standard to list
Add calibration tile to the list of stan-
dards in the instrument.
Perform diagnosis
Start the auto diagnosis procedure
here.
Green standard
Comprises the options for instrument
test.
Add docking station
Turn your spectro2go into spec-
tro2guide here or replace existing dock-
ing station.
Add standard to list
Add test standard to the list of stan-
dards in the instrument.
NOTICE
• The option
Add standard to list
is active, if the standard has not been added
yet.
• In order to delete these standards from memory use the function
Содержание 345 000 016
Страница 1: ...Measure what you see spectro2guide Pro spectro2guide spectro2go Type 345 000 016 Operating Manual...
Страница 93: ...BYK Gardner GmbH 295 024 855 E 2106 spectro2guide Pro 93 Notes...
Страница 94: ...BYK Gardner GmbH 94 spectro2guide Pro 295 024 855 E 2106...
Страница 95: ...BYK Gardner GmbH 295 024 855 E 2106 spectro2guide Pro 95...