Excell MCDSe series
Rel. 1
alsa
apparecchi medicali srl
16
Type of Error
Code
Code memorized in E
2
PROM
(in CAL mode, select “
0d
”)
2
Error on the serial communication peripheral placed on the
hand-switch control board (Alsa code: 801462)
4 slow acoustic
repetitions and then rapid
intermittence
3F
Error in the communication with the E
2
PROM
5 slow acoustic
repetitions and then rapid
intermittence
40
Absence of connection I
2
CBUS (during the use)
6 slow acoustic
repetitions and then rapid
intermittence
43
Non Master-Slave connection (during the self test)
7 slow acoustic
repetitions and then rapid
intermittence
41
Error on the Slave microcontroller
8 slow acoustic
repetitions and then rapid
intermittence
44
Errors on the Slave microcontroller
Error on a Slave microcontroller internal peripheral
Err 60
3C
Errors on the R.F. power section
Error in the internal dummy load used during the self test
Err 39
27
Error in the R.F. output power measurement circuit (with lower
setting value
Err 51
33
Error in the R.F. output power measurement circuit (with higher
setting value)
Err 52
34
Error in the bipolar power circuit with lower setting value
Err 43
2b
Error in the bipolar power circuit (with higher setting value)
Err 42
2A
Error of functioning in the monopolar circuit in the PURE mode
Err 40
28
Error of functioning in the monopolar circuit in the SPRAY
mode
Err 41
29
Error on the modulation signal of the function BLEND-1
Err 44
2C
Error on the modulation signal of the function FULG FORCED
Err 45
2d
Error on the modulation signal of the function SPRAY
Err 46
2E
Error in the reading of the output peak voltage (with lower
setting value
Err 47
2F
Error in the reading of the output peak voltage (with higher
setting value)
Err 48
30
Error in the reading of the output peak current (with lower
setting value
Err 49
31
Error in the reading of the output peak current (with higher
setting value)
Err 50
32
HF output power higher than expected
Err 98
62
Errors on the thermal protections
Thermal protection of the power supply unit
Err 27
1b
Thermal protection of the R.F. power section
Err 28
1C
2
According to the numerical indication of error which appears on the display when an anomaly occurs, the hexadecimal value
correspondent to one of the Error Codes shown in the column Code is memorized inside the E
2
PROM.