Device User’s Guide
4.
Instrument setting for absorbance / transmittance measurement
21
①
Transmission dip probe has two branches which is connected to the spectrometer and
light source.
②
Warming up the light source in the same way as example: 1.
③
Open the SpectraSmart software, open the shutter and immerse the transmission dip
probe in a container, containing solvent or solution. At this time, remove bubbles
adhering to the optical path tip and acquire the reference spectrum
(
※
)
.
④
Keep immersed, close the shutter and acquire the dark spectrum
(
※
)
.
⑤
Open the shutter, immerse the transmission dip probe in the solution to be measured,
and perform the measurement.
※
For the acquisition method, please refer to SpectraSmart User's Guide.