Alnor AXD 530 Скачать руководство пользователя страница 4

3

Using the AXD530 Safely

• When using the instrument to check air

flow in an elevated workplace, make
certain that you can safely raise and
hold the instrument while making mea-
surements. This is especially important
when you are working on a ladder.

• Avoid catching hoses or attachments in

moving machinery.

• Use the instrument only for measure-

ments using air.

• Avoid any corrosive or other danger-

ous or explosive gas mixtures.

Please dispose of used batteries in a
responsible manner.

N O T I C E

WA R N I N G

!

Содержание AXD 530

Страница 1: ...A T S I C o m p a n y OWNER S MANUAL AXD 530 MicroManometer...

Страница 2: ...tic Probe 4 Start Up Sequence 5 Zeroing the Instrument 5 The Display 7 Modes 7 AXD530 Keypad 8 The AXD530 Keypad 8 Key Functions 8 Using the AXD530 9 Basic Functions 9 To Change Units 9 Measurement Mo...

Страница 3: ...pad The AXD530 displays measurement results on a liquid crystal display LCD with additional dis play segments for units of measure and low battery indication It features two built in pressure ports fo...

Страница 4: ...instrument while making mea surements This is especially important when you are working on a ladder Avoid catching hoses or attachments in moving machinery Use the instrument only for measure ments u...

Страница 5: ...BATT above the reading From that point you will have about one hour of normal use left If comes on continuously and replaces the reading the battery voltage is too low for accurate measurements and t...

Страница 6: ...ons are tested The following will occur All the LCD segments will turn on The current software revision level will be displayed The message will appear The last unit of measurement will be displayed a...

Страница 7: ...that none of the tubing is pressurized by acciden tal squeezing Press the ZERO key until appears on the display The instrument will automatically zero its readings Repeat this procedure if necessary b...

Страница 8: ...sure will be displayed in a scientific notation format of XXEY The actual reading will be rounded to the nearest thousand XX The Y indicates the number of zeros after XX For example 12 664 is displaye...

Страница 9: ...the desired key until you see the required response on the display Key Functions ON OFF This key is used to turn the instrument on or off ZERO The AXD530 will zero the instrument any time this key is...

Страница 10: ...rement from feet per minute fpm to meters per second m s Push the UNIT key The display annunciator moves to m s selecting meters per seconds Release the key Measurement Modes You can set the AXD530 fo...

Страница 11: ...to velocity conversions Velocity fpm 4005 Pressure where Pressure is in inches H2 0 or Velocity m s 1 29 Pressure where Pressure is in Pa The velocity calculations assume standard condi tions Use Appe...

Страница 12: ...after power up Batteries should be removed from the meter if it will not be used for two 2 months or longer Calibration checks are recommended every 12 months Keep the meter clean by using a mild det...

Страница 13: ...off NOT CHANGE Then turn it on and try again Your input might simply be very stable Test your meter by pinching the hoses METER TURNS OFF The batteries are too low for your completing start up meter...

Страница 14: ..._______________________ Date of Purchase ______________________________________________________ Where Purchased ______________________________________________________ Describe Malfunction ____________...

Страница 15: ...the carrier for the damaged instru ment Call Alnor directly for assistance if necessary Repair Calibration Please follow these steps should you require factory service or repair of your Alnor Instrum...

Страница 16: ...ing instrument model number and cost of repair and or recalibration Securely package your instrument in a strong container surrounded by at least two inches 5 cm of suitable shock absorbing material I...

Страница 17: ..._____________________________________________________________ _________________________________________________________________ _________________________________________________________________ ______...

Страница 18: ...CIFICALLY AND WITHOUT LIMITATION THERE IS NO WARRANTY EXPRESSED OR IMPLIED OF MERCHANTABILITY OR FITNESS FOR PARTICU LAR PURPOSE This warranty is void if product is misused used contrary to procedures...

Отзывы: