WARNING
ALL PRESSURE ADJUSTMENTS SHALL BE MADE BY AUTHORIZED PERSONNEL ONLY.
INCORRECT PRESSURES CAN DISRUPT GAS DELIVERY TO LINE.
WARNING
THE MANIFOLD SYSTEM MUST BE LOCATED IN A WELL VENTILATED AREA. NITROUS OXIDE
CAN CAUSE UNCONSCIOUSNESS AND DEATH. OXYGEN AND N2O AND CAN GREATLY
INTENSIFY A SPARK OR FIRE. CARBON DIOXIDE AND NITROGEN
CAN CAUSE SUFFOCATION.
CAUTION
DO NOT APPLY HEAT TO DELIVERY LINE UNION WHILE IT IS CONNECTED TO THE
CONTROL UNIT. HEAT WILL HARM O-RINGS CAUSING LEAKS AND MALFUNCTIONS.
WARNING
DO NOT USE THREAD SEALANT ON ANY JOINT OR UNION. MOST SEALANTS CONTAIN
HYDROCARBONS THAT CAN BE IGNITED BY HIGH PRESSURE OXIDIZING GASES.
WARNING
KEEP HANDS AND ALL PARTS CLEAN AND FREE OF OIL, GREASE OR ANY HYDROCARBON
COMPOUND. HYDROCARBON COMPOUNDS CAN BE IGNITED BY OXIDIZING GASES.
CAUTION
DO NOT ATTEMPT TO RETROFIT THE CONTROL UNIT TO ACCOMMODATE OTHER
GASES THAT THE CONTROL UNIT WAS NOT DESIGNED FOR.
WARNING
EACH CONTROL UNIT IS DESIGNED FOR USE WITH A SPECIFIC GAS AS INDICATED ON THE
CONTROL UNIT. ANY ATTEMPTS TO ALTER ITS USE COULD RESULT IN SERIOUS
INJURIES & FAILURE OF THE CONTROL UNIT.
CAUTION
THE HEALTH CARE FACILITY MUST HAVE PROCEDURES IN PLACE TO INSURE
THAT DEPLETED CYLINDERS ARE REPLACED AS SOON AS POSSIBLE TO
ASSURE A FRESH BANK AWAITS THE NEXT BANK SHIFT.
WARNING
SUDDEN RELEASE OF HIGH PRESSURE GAS CAN CAUSE BODILY INJURY.
ALWAYS BLEED DOWN PRESSURE SLOWLY. ALWAYS WEAR SAFETY
GLASSES WHEN CHANGING CYLINDERS OR SERVICING UNIT.