Command reference
R&S
®
ZND
994
User Manual 1173.9557.02 ─ 61
Example:
*RST; :CALC:PAR:DEL 'TRC1'
Reset the analyzer and delete the (default) trace for channel no.
1.
SOUR:LPOR1 1,2
Combine the physical ports no. 1 and 2 to define the balanced
(logical) port no. 1.
LPOR1:ZDIF 27, 2
Specify a complex differential mode reference impedance of
27Ω + j * 2Ω for the defined port.
Manual operation:
See
"Single Ended Mode / Common Mode / Differential Mode"
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
<arg0>
Allows to toggle between default and renormalized reference impedance(s) for logical
port <LogPt>.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number
Parameters:
<arg0>
ON (1): Use default impedance(s)
OFF (0): Use the redefined impedances defined via
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
or via
[SENSe<Ch>:]LPORt<LogPt>:ZCOMmon
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent
logical ports
Manual operation:
See
7.3.14.11
[SENSe:]ROSCillator...
The
[SENSe:]ROSCillator...
commands control the frequency reference signal.
[SENSe:]ROSCillator:EXTernal:FREQuency
..................................................................... 994
[SENSe<Ch>:]ROSCillator[:SOURce]
.............................................................................. 995
[SENSe:]ROSCillator:EXTernal:FREQuency
<ExtClockFreq>
Specifies or queries the frequency of the external reference oscillator.
Parameters:
<ExtClockFreq>
Frequency of the external reference clock signal.
Range:
See the data sheet of your analyzer.
*RST:
10 MHz
Default unit: Hz
SCPI command reference