Command reference
R&S
®
ZND
705
User Manual 1173.9557.02 ─ 61
Example:
Suppose that the active recall set contains an active trace no. 1
and that the sweep range for a frequency sweep starts at 1 GHz.
CALC:MARK:REF ON
Create the reference marker and display it in the center of the
sweep range.
CALC:MARK:REF:X 1GHz
Set the reference marker to the beginning of the sweep range.
CALC:MARK:REF:Y?
Query the measurement value at the reference marker position.
Manual operation:
See
"Mkr <i> Stimulus / Ref Mkr Stimulus"
CALCulate<Chn>:MARKer<Mk>:REFerence:Y
<RefResponseValue>
Sets or gets the (response) value of the related reference marker.
The marker must be created before using
Setting this value is only possible in ARBitrary mode (see
. For NORMal and FIXed mode markers it is read-
only.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
This numeric suffix is ignored and may be set to any value.
Parameters:
<RefResponseValue> See
Default unit: NN
Example:
See
CALCulate<Chn>:MARKer<Mk>:REFerence:X
Manual operation:
See
"Mkr <i> Arb. Response / Ref Mkr Arb. Response"
CALCulate<Chn>:MARKer:SEARch:BFILter:RESult[:STATe]
<Boolean>
Shows or hides the bandfilter search results in the diagram area.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON - show the bandfilter search results. If no bandfilter search
has been initiated before (
BFILter
), nothing is displayed.
OFF - hide the bandfilter search results.
*RST:
OFF
Example:
See
CALCulate<Chn>:MARKer<Mk>:BWIDth
Manual operation:
See
SCPI command reference
www.allice.de