Command reference
R&S
®
ZND
660
User Manual 1173.9557.02 ─ 61
Example:
Chapter 8.2.6, "Fixture modeling"
Manual operation:
CALCulate:FMODel:ISD:COUPon:TYPE
<CouponType>
Sets/gets the coupon type to be measured for a fixture modeling with the ISD tool.
Parameters:
<CouponType>
SYMMetric2x | OPEN1x | SHORt1x | OPSHort1x
Example:
Chapter 8.2.6, "Fixture modeling"
Manual operation:
CALCulate:FMODel:ISD:DCEXtrapolat
<Boolean>
Defines whether the de-/embedding file generated by the ISD tool contains (extrapola-
ted) DC data.
Can only be set to ON, if a recent version of the ISD tool (from 2019-12 or later) is
used. ON correspons to
add_dc=1
in the tool's batch mode.
Parameters:
<Boolean>
OFF (0)
No DC data added, i.e. the touchstone data reach down to the
start frequency of the instrument sweep.
ON (1)
The ISD tool adds (extrapolated) S-parameter data for f = 0.
*RST:
OFF (0)
Manual operation:
CALCulate:FMODel:ISD:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure DUT + Fix-
ture at the active ports (see
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe]
Example:
Chapter 8.2.6, "Fixture modeling"
Usage:
Event
Manual operation:
CALCulate:FMODel:ISD:DUT:MEASure:FILename
<String>
For a fixture modeling with the ISD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Parameters:
<String>
Path to the Touchstone file, either absolute or relative to the
selected working directory (
).
SCPI command reference
www.allice.de
Allice Messtechnik GmbH