The display area shows all the S parameter measurements being made. The resulting
Touchstone file is written to C:\Users\Public\Documents\Rohde-Schwarz\Vna\Embed-
ding.
"1x Open, 1x Short" (ISD only) the measurement proceeds in two
steps: one for the "1x Open" and one for the "1x Short" coupon. The label of the "Mea-
sure" button changes accordingly.
Remote command:
CALCulate:FMODel:ISD:COUPon:MEASure
CALCulate:FMODel:ISD:COUPon:MEASure:OPEN
CALCulate:FMODel:ISD:COUPon:MEASure:SHORt
CALCulate:FMODel:SFD:COUPon:MEASure
CALCulate:FMODel:EZD:COUPon:MEASure
Load File / 1x Open Preset / 1x Short Preset
Allows you to load the coupon properties from a Touchstone file (
*.s*p”
).
Remote command:
CALCulate:FMODel:ISD:COUPon:MEASure:FILename
CALCulate:FMODel:ISD:COUPon:MEASure:FILename:CLEar
CALCulate:FMODel:ISD:COUPon:MEASure:OPEN:FILename
CALCulate:FMODel:ISD:COUPon:MEASure:SHORt:FILename
CALCulate:FMODel:SFD:COUPon:MEASure:FILename
CALCulate:FMODel:SFD:COUPon:MEASure:FILename:CLEar
CALCulate:FMODel:EZD:COUPon:MEASure:FILename
CALCulate:FMODel:EZD:DUT:MEASure:FILename:CLEar
Active
Before starting to
the test coupon, use these checkboxes to indicate the
physical ports to which it is connected.
Table 5-5: Allowed numbers of active ports
"Sym 2x Thru"
"1x Open, 1x Short"
other
Single-ended
2
1 or 2
1
Balanced
2 or 4
2 or 4
2
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:COUPon[:STATe]
CALCulate:FMODel:SFD<Ph_pt>:COUPon[:STATe]
CALCulate:FMODel:EZD<Ph_pt>:COUPon[:STATe]
5.13.5.3
Measure DUT + Test Fixture
The "Measure DUT + Test Fixture" section allows you to measure the whole structure,
i.e. the DUT with the test fixture.
For EZD this section is only available if
"Advanced Settings".
Offset Embed softtool
www.allice.de
Allice Messtechnik GmbH