GUI reference
R&S
®
ZND
314
User Manual 1173.9557.02 ─ 61
Field Separator
For ASCII (
*.csv
) and Matlab (
*.dat
) export only: defines the separator the analyzer
uses to separate different numbers in each line of the file.
Decimal Separator
For ASCII (
*.csv
) files only: Selects either the "Point" or the "Comma" (if needed to
process the exported data with an external application) as a separator for decimal
numbers.
Save
Stores the trace data, according to the selected options.
Tip:
Note the conditions described in
"Conditions for Touchstone file export"
Remote command:
MMEMory:STORe:TRACe:PORTs
MMEMory:STORe:TRACe
MMEMory:STORe:TRACe:CHANnel
Comment Added to File
Defines a comment to be added to the exported trace data file.
If empty (←"Delete"), no comment is added.
Remote command:
MMEMory:STORe:TRACe:OPTion:COMMent
5.5.11.3
Select Parameter dialog
The "Select Parameter" dialog provides a selection of measurement results (e.g. S-
parameters) or traces, e.g. for trace import, import of power correction coefficients, limit
line import.
Access:
The dialog may be called from several dialogs, for example on pressing
"Open" in the
Select All / Deselect All
During trace data import, selects/deselects all traces contained in the opened trace file.
Auto Distribute
Available for trace data import only.
If checked, a selected trace S
ij
is imported as a memory trace for all data traces in the
current recall set that are measuring S
ij
.
Trace Config softtool
www.allice.de
Allice Messtechnik GmbH