Command reference
R&S
®
ZND
1063
User Manual 1173.9557.02 ─ 61
The configuration is valid if
●
every VNA port is either connected to a matrix port or directly assigned to a DUT
test port
●
every matrix that is part of the configuration is connected to the VNA by at least
one RF connection
●
every VNA port that is not connected to a matrix is assigned to a test port
●
the test ports are numbered consecutively, starting with 1
Example:
SYSTem:COMMunicate:RDEVice:SMATrix:
Usage:
Event
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:STARt
Starts a manual RF connection configuration.
Note:
a manual RF connection configuration must be performed within a transaction,
SYSTem:COMMunicate:RDEVice:SMATrix:
SYSTem:COMMunicate:RDEVice:SMATrix:CONFigure:
. The modified configuration will not be applied until the end of this transaction.
An RF connection configuration comprises the RF connections between the VNA, a set
of switch matrices and the DUT test ports.
At transaction start the R&S
ZND creates an in-memory copy of the active configura-
tion and strips off all RF conncections from this copy. Subsequent RF connection
actions such as
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:CONFigure:
SYSTem:COMMunicate:RDEVice:SMATrix<Matr>:CONFigure:MTESt
operate on the copy.
However, the copy contains the very same switch matrices as the original, so the ade-
quate set of matrices has to be selected beforehand (using manual control or the
remote command
).
If a previous transaction is active, it will be silently rolled back.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH