Command reference
R&S
®
ZND
1029
User Manual 1173.9557.02 ─ 61
SOURce:POWer:CORRection:TCOefficient:INSert<ListNo>
<Frequency>,
<TransCoeff>[, <DeEmbedding Position>]
Inserts a new frequency and transmission coefficient at position no. <ListNo> in the
transmission coefficient list. The following points are shifted down in the list.
Suffix:
<ListNo>
.
List entry no.
Parameters:
<Frequency>
Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range:
Stimulus values outside the frequency range of the
analyzer are allowed.
*RST:
n/a
Default unit: Hz
<TransCoeff>
Transmission coefficient
Range:
-300 dB to +200 dB
*RST:
n/a
Default unit: dB
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using
SOURce<Ch>:POWer<PhyPt>:CORRection:
.
Example:
SOURce:POWer:CORRection:TCOefficient[:STATe]
Manual operation:
"Frequency / Transm. Coefficients: Insert, Delete, Delete
SOURce:POWer:CORRection:TCOefficient[:STATe]
<Boolean>[, <DeEmbedding
Position>]
Enables or disables the use of two-port transmission coefficients for power correction.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH