Command reference
R&S
®
ZND
1021
User Manual 1173.9557.02 ─ 61
Note
●
Port groups must be created in ascending order, i.e. first create port group 1, then
port group 2 etc.
●
By default there is a single port group, which contains all physical VNA ports
("default port group").
●
Redefinition of an existing port group is not allowed (exception: "default port
group"). Use
to dissolve existing port groups
first.
●
It is not allowed to assign the same physical VNA port to multiple port groups.
●
When a port group is created, the number of ports of the corresponding DUT is set
to the port group size and "aligned" port connections are etablished.
Suffix:
<Ch>
.
Channel number
<Grp>
Port group number
Parameters:
<pport1>,<pport2>... Physical VNA port numbers
7.3.15.3
SOURce:POWer:CORRection:COLLect...
The
SOURce:POWer:CORRection:COLLect...
commands allow you to set up a
scalar source power calibration.
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt]
<NoReadings>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels and ports.
SOURce:POWer:CORRection:NREadings
.
Parameters:
<NoReadings>
Number of readings
Range:
1 to 100
*RST:
2
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance
<Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.