Access the service menu:
“SERVICE”
•
Enter carefully in the top left box the temperature read on the reference thermometer during the
second step
•
Enter carefully in the top right box the temperature read on the reference thermometer during
the first step
•
Enter carefully in the bottom left box the temperature read on the reference thermometer during
the second step
•
Enter carefully in the bottom right box the temperature read on the reference thermometer during
the first step.
•
Make sure the data entered are correct and click on validate (green button)
The temperature on the main screen will adjust according to the calibration.