TECHNICAL FEATURES
Jet nebuliser bowl
Maximum fill:
8 ml
Aerosol output:
0.45 ml/min Approx.
Particle size (MMAD)*: 2.1 microns
Manufactured from non toxic material
* The value shown is conservative and in certain conditions may be surpassed
Microbial Contamination
The jet nebuliser bowl and its accessories are for single person use only. If used by other persons cross
infection may occur.
Interactions
The materials used that come into contact with medication have been tested with a wide range of
medication in common use such as ambroxol hydrochloride, ipratropium bromide, bromhexine
hydrochloride, beclomethasone diproprionate, acetylcysteine and salbutamol. However, due to the
variety and continuous evolution of the medication, the possibility of a chemical interaction cannot be
excluded. Once opened, the medication should be used as soon as possible and avoid prolonged
contact between medication and nebuliser.
Disposal
To safeguard the environment, when the appliance has reached the end of its useful life take it
to an authorised electric appliance disposal centre where the raw materials can be recycled.
Please check to see if this is applicable in your area.
TROUBLE SHOOTING
If the
Ventalair
®
MAX
fails to operate on 230 volt power:
• Check the electric socket with another appliance or try another electric socket
• Check electric cable for damage
• The
Ventalair
®
MAX
is fitted with a thermal fuse which cannot be checked and is not replaceable
• If still not operational return for service
If the jet nebuliser bowl is misting poorly or is taking too long to nebulise your medication:
• Replace your jet nebuliser bowl every 6 months if it is used regularly.
• Check that the centre jet (8b) is in place.
• Check that the hole in the jet (8b) is not blocked with dry medication. Soak the jet nebuliser bowl in
water and try again. Do not tamper with the holes, as this will alter the particle size of your medication.
• Switch pump on and off to ensure air is flowing from the air outlet port (4)
• Check inlet filter (3) for blockage and change if the filter is discoloured
• Check the tubing for leaks, cracks or kinking
• If moisture collects inside air tubing when nebulising medication, disconnect jet nebuliser bowl on
completion, operate pump to pass air through the tube until dry
• Once a year take your nebuliser system to your pharmacy to be checked
CAUTION: Do not dismantle the
Ventalair
®
MAX
under any circumstances as this may void your warranty.
Refer servicing to authorised service centre. The
Ventalair
®
Max
has an oil free compressor and requires
no servicing. Do not oil.
UNIT TECHNICAL FEATURES
Mod. Ventalair
®
MAX
Voltage
230V~ 50Hz 140VA
Fuse
T 2.5A – 250V
Continuous use
Air flow
10 l/min
Dynamic air flow:
6.2 l/min
Dimensions:
180(L) x 300(D) x 120(H) mm
Weight:
1.6 Kg
Manufactured to comply with Medical Directive:
93/42 EECand STD. EN 60 601-1
Operating requirements:
Temperature min. 10
°
C; max 40
°
C
Air humidity min. 10%; max 95%
RH
Storage requirements:
Temperature min. -25
°
C; max 70
°
C
Air humidity min. 10%; max 95%
RH
Operating/storage atmospheric pressure: min. 690 hPa; max. 1060 hPa
P