Compare the control solution test result to the control solution range printed
on the test strip vial. The test result should fall within this range. Each vial of
test strips may have a di
ff
erent control solution range. If the test result you
get is not within this range, the meter and test strips may not be working
properly. Repeat the control solution test. Refer to the GlucoDr SuperSensor
Test Strip Instruction for the use.
Compare Control Solution Test Results
The Control Solution range printed on the test strip vial is for the GlucoDr
SuperSensor Control Solution only. It is used to check the performance of your
meter and test strips. It is NOT a recommended range for your blood glucose level.
Out-of-range test results may be caused by one or more of the followings:
- Expired or contaminated control solution,
- Expired or damaged test strip,
- Use of control solution or test strip past its discarding date,
- Error in performing the test,
- Failure to shake the control solution vial well,
- Code number printed on the test strip vial does not match the one displayed on
the meter,
- Meter, test strips or control solution are kept too warm or too cold,
- or Meter malfunction