7.  To test that the address has been accepted, type the new address followed by the 

Enter key. The computer should respond with the new address followed by one line 

of data.

8.  Put some sort of external label on the device that denotes its address. If you forget 

the device’s address, you will have to check the device individually (with all other 

units disconnected).

9. 

Remove the device from the Multi-Drop Box and connect the next device to be 

addressed. Repeat until all of the units have a unique address.

Daisy Chaining Multi-Drop Boxes For More Devices

Up to 26 flow and/or pressure devices can be attached to a single RS-485 port by cabling 

up to four Multi-Drop Boxes together. Valid RS-485 addresses for our flow and pressure 

instruments are the 26 letters of the alphabet, A-Z. 
To daisy chain Multi-Drop Boxes when only 

metering 

and/or 

gauging

 devices are being 

used, simply use a double ended DC-62 type 8 pin mini-DIN (or 6 pin Locking Industrial) 

cable to connect any port on one box to any port on another.

To daisy chain Multi-Drop Boxes when 

controlling

 devices are being used, use a double 

ended DC-62 type 8 pin mini-DIN (or 6 pin Locking Industrial) cable to connect any port 

on one box to any port on another. 

It is also necessary to daisy chain additional wires of 

sufficient gauge (to power the extra units) between the power and ground sockets on the 

terminal blocks

 of the Multi-Drop Boxes. 

CAUTION: Failure to provide proper wire gauge for the expected load between 

boxes may result in damage and possible fire!