AUF610 Series Operation Manual 

                                                                                                                 

OP610.1.1.5R1ENG

           

 

23

 

M22 

Window for entering the viscosity of the non-standard liquids 

M23 

Window for selecting the proper transducers 
There are 14 different types of transducers for selection. 
If the user-type-transducers are used, 4 user type wedge parameters, which will be 
prompted by the software, should be entered following. 
If the 

π

  type transducers are used, 3

π

  type transducers and pipe parameters should 

be entered following. 

Model          S             M              L              HS            HM 

Transducer      7 Clamp-on          7.Clamp-on      15. Clamp-on                  12                            8 

Type           TS-1            TM-1        TL-1          Standard-S1    Standard-M1

M24 

Window for selecting the transducer mounting methods, Four methods can be selected:
(0) V - Method   (1) Z - Method   (2) N  - Method   (3) W - Method 

M25 

Display the transducer mounting spacing 

M26 

Entry to store the parameter configuration into the internal NVRAM 

M27 

Entry to load one set of saved parameters 

M28 

Select YES or NO for the instrument to determine whether or not to hold ( or to keep ) 
the last correct value when poor signal condition occurs. YES is the default setup 

M29 

Enter a value ranging from 000 to 999.    0 is the default value 

M30 

Window for selecting unit system. Default value is ‘Metric’. The change from English to 
Metric or vice versa will not affect the unit for totalizers. 

M31 

Window for selecting flow rate that will be used by the instrument afterward. Flow rate 
can be in 
0.Cubic                                    (m3)                    5.Cubic feet                            (cf) 
1.Liter                                      ( I )                      6.USA liquid barrel                (bal) 
2.USA  gallon            (gal)         7.Imperial Liquid barrel     (ib) 
3.Imperial Gallon                    (igl)                    8.Oil barrel                              (ob) 
4.Million USA gallon       (mgl) 
The flow unit time base can be per day, per hour, per minute or per second 

M32 

Window for selecting the totalizer’ working unit 

M33 

Select totalizer multiplier 

M34 

Turn on or turn off the NET totalizer 

M35 

Turn on or turn off the POS totalizer 

M36 

Turn on or turn off the NEG totalizer 

M37 Totalizer 

reset 

Restore the instrument to the default parameters as the manufacturer did by pressing 
the dot key followed by the backspace key. CAUTION: Please note the current 
parameters before doing restoration as this cannot be undone. 

M38 

Press-key-to-run or to stop totalizer for easier calibration 

M39 

Operational interface language selection in Chinese and English. This selection makes 
it possible that more than 2 billions of people on the word can read the menu. 

M40 

Flow rate damper for a stable value. The input range is 0 to 999 seconds. 

M41 

Lower flow rate cut-off to avoid invalid accumulation. 

M42 

Zero point setup under the condition when there is no liquid running inside the pipe. 

Содержание AUF610 Series

Страница 1: ...Transit Time Ultrasonic Flowmeter Model AUF610 Series Operation Manual Alia Group Inc URL http www alia inc com 113 Barksdale Professional Center Newark DE 19711 USA E mail alia alia inc com TEL 1 30...

Страница 2: ...9 2 V method Installation 16 2 9 3 Z method Installation 16 2 9 4 W method Installation 16 2 10 Installation Verification 17 2 10 1 Signal Strength 17 2 10 2 Signal Quality 17 2 10 3 Total Transit Ti...

Страница 3: ...e the Working Timer 21 3 25 How to use the manual totalizer 21 3 26 How to check the ESN and other details 21 3 27 How to know how long the battery will last 21 3 28 How to charge the built in battery...

Страница 4: ...e rich software provides a user friendly interface that provides the operator with access to more information for accurately monitoring and analyzing pipe flows The AUF610 uses a patented balanced low...

Страница 5: ...called the Z method If the sound transverses the pipe twice it is called the V method If the sound transverses the pipe four times then it is known as the W method The flow meter operates by alternate...

Страница 6: ...itten in the software and can only be accessed by the manufacturer In the case of an equipment failure this number located on 0menu window M61 MUST be provided when contacting the manufacturer 1 5 Par...

Страница 7: ...610 Series Operation Manual OP610 1 1 5R1ENG 7 AUF 610 HANDHELD ULTRASONIC FLOWMETER AUF610 POS 2 m3 Flow 0 0000 m3 h Vel 0 0000 m s S 000 000 Q 00 1 Front view LCD DISPLAY LED Charging Indicator Keyp...

Страница 8: ...tion Manual OP610 1 1 5R1ENG 8 Standard Type Temperature Sensor 15mm 6000mm High Temperature Type Sensor 15mm 1000mm Transducer Standard Cables 5M Blue Terminal 5m Blue Terminal Cable 5m 2 Red T ermi...

Страница 9: ...cals reclaimed water cooling water river water plant effluent etc The instrument transducers are mounted outside the pipe and are therefore not in contact with the process and have no moving parts The...

Страница 10: ...user configurable Velocity 32m s Pipe Size 15mm 6000mm Rate Units Meter Feet Cubic Meter Liter Cubic Feet USA Gallon Imperial Gallon Oil Barrel USA Liquid Barrel Imperial Liquid Barrel Million USA Ga...

Страница 11: ...e OFF to turn off the power Once the flow meter is switched on it will run a self diagnostic program checking first the hardware and then the software integrity If there are any issues an error messag...

Страница 12: ...zeroing calibration and modification password setup M50 M53 items are for the built in logger M60 M78 items are for time keeper initialization version and ESN information viewing and alarms M82 is fo...

Страница 13: ...FF or ON Totalizers M34 35 36 Positive Negative Net Totalizers Totalizer reset see Item M37 page 22 Now if the transducers are installed correctly and the above is done the meter is ready to measure P...

Страница 14: ...ystem about to be measured The optimum location will be length of straight pipe that is always full of process and has no bubbles The piping can be in vertical flow going up or horizontal position but...

Страница 15: ...of the transducers are factors critical for the specified measurement performance of the system Great care should be taken when installing the transducers Here are some guidelines Locate an optimum po...

Страница 16: ...alled reflective mode or single bounce method 2 9 3 Z method Installation Z method is commonly used when the pipe diameter is between200millimeters and 6000 millimeters 2 9 4 W method Installation not...

Страница 17: ...gnal Quality Signal quality is indicated on the lower right hand side of the display and denoted as Q The higher a Q value the better the guarantee of a reliable measurement Under normal pipe conditio...

Страница 18: ...t a proper totalizer Make sure that the totalizer pulse rate is appropriate for the application the receiving instrument yet still output the resolution necessary to properly characterize the flow If...

Страница 19: ...ssword input prompt displays If the password is forgotten please contact the factory 3 14 How to use the built in data logger The data logger has a space of 24K bytes of memory which will hold about 2...

Страница 20: ...g 1 Alarms on when there is no receiving signal 2 Alarms on when there is poor signal received 3 Alarms on when the flow meter is not in normal measurement modes 4 Alarms on reverse flow 5 Alarms on t...

Страница 21: ...monthly totalizer and a yearly totalizer 3 24 How to use the Working Timer Use the working timer to check the time that has passed with a certain kind of operation For example use it as a timer to sho...

Страница 22: ...dow for entering pipe wall thickness M13 Window for entering the inner diameter of the pipe M14 Window for selecting pipe material Standard pipe materials that the user need not know the speed include...

Страница 23: ...is Metric The change from English to Metric or vice versa will not affect the unit for totalizers M31 Window for selecting flow rate that will be used by the instrument afterward Flow rate can be in 0...

Страница 24: ...r viewer It functions as a file editor Use Dot backspace UP and DN keys to browse the buffer M54 Not used M55 Nod used M56 Not used M57 Not used M58 Not used M59 Not used M60 Settable 99 year calendar...

Страница 25: ...alue has an obvious difference with the actual fluid sound speed the pipe parameters entered and the transducer then installation should be checked again M93 Displays total transit time and delta time...

Страница 26: ...problems Contact the factory 5 2 Operational Error Codes and Corrective Actions The AUF610 series ultrasonic flow meter will display a Code in the lower right corner with a single letter like I R etc...

Страница 27: ...le Zero Offset value incorrectly entered in M44 Enter 0 in M44 Problem with transducer installation There is a Zero Point Try to zero the instrument by using M42 and make sure that the flow inside the...

Страница 28: ...ddddE dd CR LF DQM CR Return flow rate per minute d ddddddE dd CR LF DQS CR Return flow rate per second d ddddddE dd CR LF DV CR Return flow velocity d ddddddE dd CR LF DI CR Return POS totalizer dddd...

Страница 29: ...PDI CR would return 1234567E 0m3 F7 CR LF acts as the starter of check sum which is yielded by adding up the string 2BH 31H 32H 33H 34H 35H 36H 37H 45H 2BH 30H 6DH 33H 20H Please note that there will...

Страница 30: ...trument by transmitting the M command along with the keypad code Using this function remote operation of this instrument can be realized even via the Internet KEYPAD KEY HEX CODE KEY DECIMAL CODE KEY...

Страница 31: ...e applied standard Measurement Uncertainty Measurement uncertainty is the statistical calculation of the total error exhibited by a measuring instrument as a result of a series of measurements taken a...

Страница 32: ...r more the instrument has many troubleshooting features and measurements that allows the user to track down issues that may be affecting optimum performance Pointers to Be Aware of Accurately determin...

Страница 33: ...065 353 386 8 8 625 0 322 7 981 539 581 10 10 75 0 365 10 02 755 824 12 12 75 0 406 11 938 1001 1089 14 14 0 438 13 124 1180 16 16 0 5 15 1543 PVC and CPVC Pipes Schedule 80 Nominal Minimum Nominal We...

Страница 34: ...0 358 0 599 1 1 4 1 66 5S 0 065 1 53 10S 0 109 1 442 STD 40 40S 0 14 1 38 XS 80 80S 0 191 1 278 160 0 25 1 16 XXS 0 382 0 896 1 1 2 1 9 5S 0 065 1 77 10S 0 109 1 682 STD 40 40S 0 145 1 61 XS 80 80S 0...

Страница 35: ...237 4 026 XS 80 80S 0 337 3 826 120 0 438 3 624 160 0 531 3 438 XXS 0 674 3 152 5 5 563 5S 0 109 5 345 10S 0 134 5 295 STD 40 40S 0 258 5 047 XS 80 80S 0 375 4 813 120 0 5 4 563 160 0 625 4 313 XXS 0...

Страница 36: ...2 XS 60 80S 0 5 9 75 80 0 594 9 562 100 0 719 9 312 120 0 844 9 062 140 1 8 75 160 1 125 8 5 12 12 75 5S 0 156 12 438 10S 0 18 12 39 20 0 25 12 25 30 0 33 12 09 STD 40S 0 375 12 40 0 406 11 938 XS 80S...

Страница 37: ...nzene 1298 Carbon disulfide 1149 Carbon tetrachloride 926 Castor oil 1474 Chloroform 995 Ether 985 Ethylene glycol 1644 Glycerol glycerine 1904 Heptane 1138 Hexane 1203 Kerosene 1324 Mercury 1450 Meth...

Отзывы: