Alfaa EcoStream Lite Скачать руководство пользователя страница 5

 

MAINTENANCE

 

The basic unit is designed to operate with minimal maintenance requirements providing the 
minimum water characteristics are met as are outlined in the section “Application Guidelines”. 
However there are two regular maintenance requirements common to all UV systems: quartz 
cleaning and lamp replacement. 

Cleaning 

Unless using deinozed or RO permeate water, minerals in the water will eventually coat the quartz 
sleeve (which protects the lamp). This coating must be cleaned off periodically because it reduces 
the amount of UV light reaching the water, thereby reducing disinfection performance. 

Periodically, check the sleeve and clean it if you can see a mineral coating starting to form. If sleeve 
requires cleaning, refer to below instructions but re-install the original lamp.  

Quartz Sleeve Replacement/Cleaning 

1.

 

If the lamp is in the system, remove the lamp and carefully set it aside as described in the lamp 
installation/replacement section.  

2.

 

Shut off the upstream water supply that feeds water into the reactor chamber. Depressurize and 
drain the system by disconnecting the inlet/outlet from the reactor chamber. 

3.

 

Unscrew and remove the QG sealing nut from the top of the reactor (Figure 1). Make sure to 
remove the QG sealing O-ring and keep it carefully (Figure 2). 

4.

 

Carefully slide the sleeve out of chamber (Figure 3). In case it is initially tight, gently try rotating 
the QG while also pulling it out. Also make sure that the sleeve is 

not at an angle

 as otherwise 

pressure will be applied on the sides of the sleeve and against the reactor chamber causing the 
sleeve to fracture. 

5.

 

Clean the quartz sleeve, or replace them with a new one. To clean the sleeve, use a mild acid 
solution such as 10% Citric Acid or household cleaners such as vinegar. 

6.

 

Reinstall the quartz sleeve in reverse order. 

Carefully

 slide the quartz sleeve into the reactor 

through the QG socket until it is locked into place in the internal QG holder. Ensure that the 
sleeve is inserted straight and 

not at an angle

 as doing so will put pressure on the wall of the 

sleeve and can cause it to crack. Install the quartz sealing o-ring [7] onto the sleeve until it rests 
against the QG socket. 

7.

 

Reinstall the QG socket compression nuts by turning clockwise. This nut should be hand 
tightened only. 

8.

 

Slowly turn on the water and pressurize the reactor to verify that there are no leaks. 

9.

 

Reinstall the lamp as described in the lamp installation/replacement section and reconnect all 
the electrical connections to ensure that the system is operating properly. 

 

 

Figure 1 

 

Figure 2 

 

Figure 3 

 

Содержание EcoStream Lite

Страница 1: ...OPERATIONS AND MAINTENANCE MANUAL EcoStream Lite...

Страница 2: ...not approved recommended or sold by the manufacturer dealer may cause an unsafe condition 7 Intended for indoor use only Do not install this UV disinfection system where it will be exposed to the weat...

Страница 3: ...is affected by many factors and the following should be looked at prior to the installation of the UV system 1 UV Transmission transmittance deals with the effectiveness in which the 254 nanometer wa...

Страница 4: ...n on the inlet valve and SLOWLY fill the reactor chamber with water Once the chamber is full check for leaks from the fittings If a leak is evident drain system and reconnect the fittings 5 In case th...

Страница 5: ...f the reactor Figure 1 Make sure to remove the QG sealing O ring and keep it carefully Figure 2 4 Carefully slide the sleeve out of chamber Figure 3 In case it is initially tight gently try rotating t...

Страница 6: ...try twisting the connector as it will break Pull only on the lamp pins connector Wipe off any oils with alcohol and a soft cloth Warning Depending on when the system was powered down the UV lamp migh...

Страница 7: ...ull and void if unauthorized repairs alteration or modification is made to be product 4 The serial number must not be defaced or altered in any way 5 The Warranty extended herein is in lieu of all imp...

Страница 8: ...l W Mumbai 400013 India Tel 91 22 66612300 Fax 91 22 66612333 E Mail indsales alfauv com Website www alfauv com Service Office 307 A2 Shah Nahar Industrial Estate Lower Parel W Mumbai 400013 India Tel...

Отзывы: