6
Technical data
Study the instructions carefully. The items refer to the Parts list and service kits section. Handle scrap correctly.
Man = Manually activated
NC = Normally closed.
A/A = Air/air activated.
6.1
Technical data
Design temperature
Sterilisation In Place (SIP)
Max. 135
°
C / (275
°
F) (at 1 hour)
Autoclave clearing
Pneumatic actuator can not be autoclaved.
Manual actuator can be autoclaved.
Autoclave (manual valves only can be autoclaved)
Max. 130
°
C (266
°
F) (at 1 hour)
Design pressure
Minimum working pressure
Full vacuum
Maximum working pressure
7 bar (101 PSI)
Other design data
pH range
3-11
Viscocity
0-1000 cP
Material
Housing (valve body)
AISI 316L
Actuator cover
AISI 304
Diaphragm
Silicone
According to FDA specification of approved material (FDA 21 CFR §
177.2600)
EPDM
According to FDA specification of approved material (FDA 21 CFR §
177.2600)
See section 5.3 for information about the diaphragms
Surface treatment
Internally
High grade polished Ra 0.5
µ
m
or (SFI) electro polish Ra 0.4
µ
m (SF4)
Externally
Ra 0.8
µ
m
Operating data for pneumatic actuator
Actuator function:
SA: Pneumatic upward movement, spring return (NC)
AA: Pneumatic upward and downward movement
Man: Manually operated
Operating data: Pneumatic actuator (Spring operated) SA
Control air
Dry, free from particies and oil (ISO 8573.1 Class 2.2.1)
Supply pressure, recommended
6 bar (88 PSI)
Supply pressure, minimum
5.5 bar (79 PSI)
Supply pressure, maximum
7 bar (101 PSI)
Operating data - Pneumatic actuator (Air operated) AA
Control air
Dry, free from particies and oil (ISO 8573.1 Class 2.2.1)
Supply pressure, recommended
3.5 bar (51 PSI)
Supply pressure, minimum
3 bar (44 PSI)
Supply pressure, maximum
4 bar (58 PSI)
Air consumption - Pneumatic actuators (AA & SA)
Air consumption RDV-UP ½”
0.03 NI/stroke at 4 bar
Air consumption RDV-UP 1”
0.12 NI/stroke at 4 bar
Air consumption RDV-UP 1 ½”
0.3 NI/stroke at 4 bar
Air connection
M5 thread
Hose, quick connection
4mm hose
28