Operator Control

Jog shuttle
PUSH:  SW version
TURN:  Select menu

PUSH:  Config menu

TURN:  Select option
 

Left: Decrease

 

Right: Increase

LCDisplay

999.9 mS/cm

139.9 °C

0.001 mS/cm

000.1 °C

Max. indication

Min. indication

The back-lit display indicates the conductivity in mS/cm and the 

temperature in °C. The jog shuttle is used for simple configuration of 

the measuring range(s) and the temperature coefficient(s) as well as 

the temperature range.

Refer to the detailed description under “Configuring the Conductivity 

Sensor”. 

Configuring the Alfa Laval Conductivity Sensor

Jog-shuttle

Step

Subject

Range

Display

Turn

Push

Note

1

Power on

starting 

...

2

Actual value

 

1

yyyyy  mS 

zzzzz  °C

Push

3

SW version

Actual version

Push

4

Actual value

 

1

yyyyy  mS 

zzzzz  °C

Right

5

Set Range 1

1

4...20 mA 

Actual range, mS

Push

Note 1

6



4...20 mA 

Set range, mS

Left/Right

Push to set

Left: Decrease 

Right: Increase

7

1

4...20 mA 

New range, mS

Right

8


1

Resolution 

Actual Coefficient

Push

Note 2

9



Resolution 

Set coefficient

Left/Right

Push to set

10


1

Resolution 

New coefficient

Right

11

Set Range 2

2

4...20 mA 

Actual range, mS

Push

Repeat steps 5...10 for ranges 2...4.

12

Set temperature 

range

4...20 mA     °C 

Actual range

Push

13

Set temperature 

range

4...20 mA     °C 

Select range

Left/right

Push to set

Note 3

Содержание TE67K Series

Страница 1: ...Alfa Laval Conductivity Transmitter TE67Kxxxxxxxxx have according to the type been developed for applications where conduc tivity and temperature are to be measured Inductive conductivity measurement is based on the principle of two series connected toroidal core transformers 1 and 2 The primary side of the transformer 1 is controlled by an AC voltage generator The liquid flowing through the chann...

Страница 2: ... Insert length L 40 Standard sensor shaft Insert length L 87 Extended sensor shaft Mechanical Installation Dimensional Drawing After Installation and Configuration Check the leak tightness of the sleeve Check the tightness of glands or M12 plugs Check the tightness of the cover ø89 Clamp 1 1 2 ISO2852 DIN32676 TEK67K1311111xx TE67K1211111xx Clamp 2 ISO2852 DIN32676 TEK67K1711111xx TE67K1811111xx G...

Страница 3: ...our measuring ranges All four measuring ranges can individually be set to a range according to the table The 3 A mark is valid only when the product is mounted in a 3 A marked counter part and installed according to the installation manual Use also a 3 A marked O ring or gasket if relevant The 3 A marked products conforms to the 3 A Sanitary Standard criteria Materials and surfaces fulfill the FDA...

Страница 4: ...y Sensor Configuring the Alfa Laval Conductivity Sensor Jog shuttle Step Subject Range Display Turn Push Note 1 Power on starting 2 Actual value 1 yyyyy mS zzzzz C Push 3 SW version Actual version Push 4 Actual value 1 yyyyy mS zzzzz C Right 5 Set Range 1 1 4 20 mA Actual range mS Push Note 1 6 1 4 20 mA Set range mS Left Right Push to set Left Decrease Right Increase 7 1 4 20 mA New range mS Righ...

Страница 5: ...ns contained in this manual were effective at the date of printing However as continuous improvements are our policy we reserve the right to alter or modify any unit specification on any product without prior notice or any obligation Reservation and contact infomation 2012 Alfa Laval Corporate AB This document and its contents is owned by Alfa Laval Corporate AB and protected by laws governing int...

Страница 6: ...lenght 37 mm TE67Kx7xxxxxxx Clamp DN 50 ISO2852 Clamp DN 51 DIN32676 with sensor shaft lenght 84 mm TE67Kx8xxxxxxx DN 32 Din11851 L 37mm TE67KxAxxxxxxx DN 40 Din11851 L 37mm TE67KxBxxxxxxx DN 50 Din11851 L 37mm TE67KxCxxxxxxx DN 32 Din11851 L 84mm TE67KxDxxxxxxx DN 40 Din11851 L 84mm TE67KxExxxxxxx DN 50 Din11851 L 84mm TE67KxFxxxxxxx Surface finish wetted parts Ra 0 8µ TE67Kxx1xxxxxx Ra 0 4µ Elec...

Страница 7: ...UGH December 31 2012 Timothy R Rugh Executive Director 3 A Sanitary Standards Inc The issuance of this authorization for the use of the 3 A Symbol is based upon the voluntary certification by the applicant for it that the equipment listed above complies fully with the 3 A Sanitary Standards designated Legal responsibility for compliance is solely that of the holder of this Certificate of Authoriza...

Страница 8: ...Installation Manual 5850 023 TE91K022 EN3 BADK 082012 ...

Отзывы: