ALCOLOCK L OEM

4

CALIBRATING A HANDSET

1. 

Power the handset on by holding down either the right or left button.

2. 

Access the Menu by pressing and holding the left button. Scroll through the menu and select 

System Maintenance

. The four digit code which is provided with the L OEM accesses this 

menu.

3. 

Enter the code and move the cursor to 

OK

? when finished. Push the right button to Select. 

This enters the Service menu.

4. 

Scroll through the Service menu and press select 

Calibration

.

5. 

The current handset Date/Time is displayed. To proceed, select OK with the right button. This 

enters the Select Type menu.

NOTE! If a tachograph is installed the date/time cannot be changed.

6. 

The Select type menu contains the calibration options. Scroll to 

Dry Gas

 and push the right 

button to Select. This enters the Select value menu.

7. 

The Select value menu contains the gas standard concentration options. Verify that 130PPM 

is on the canister label, and use the left button to scroll to 

130PPM

. Push the right button to 

Select. This enters the Altitude menu.

8. 

The Altitude menu contains the altitude correction factor. Verify the altitude value using the 

altitude reference guide, which is on the last page of this guide. The left button increases in 

increments of 

200

 meters. Press the right button to Accept.

9. 

Wait

 message is displayed followed by a 2:00 minute warm-up. The handset is ready for 

calibration when the 

TURN ON GAS

 message is displayed.

10. 

Insert the delivery tube from the regulator valve into the mouthpiece port. Keep holding the 

handset as the delivery tube is rigid.

11. 

Press the valve down to release the gas while keeping 

hold of the handset (see figure). The handset analyzes 

the gas standard and displays a 

Blowing…

 message 

followed by an 

Analyzing

 message.

Continue holding the valve down until the handset clicks 

and then beeps, indicating that the test is complete.

12. 

The handset analyzes the test, which is followed by a 

Wait 

message and 0:45 second countdown. The handset 

is ready for the second sample when the 

TURN ON GAS

 

message is displayed.
Depending on the sensor calibration drift, up to 4 samples may be required.

•  If only 2 tests are required, a 

Verification OK

 message is displayed when thecalibration 

is complete

•  If 3 or 4 tests are required, a 

Calibration OK

 message is displayed when calibration is 

complete

13. 

Press the right button to Finish (there is also the option to retry by pressing the left button).

Calibration is complete. 
The L OEM handset returns to ready state and 

Blow for 5 seconds

 is displayed.

NOTE! If the calibration fails on numerous attempts, the sensor may need 

replacement. Contact ACS for return instructions.

Содержание L

Страница 1: ...Service Guide ...

Страница 2: ...s International Inc and are used under license Alcohol Countermeasure Systems is the trading style of Alcohol Countermeasure Systems Corp 2013 Alcohol Countermeasure Systems The information disclosed in this document is the valuable property of Alcohol Countermeasure Systems and all copyright and other proprietary rights to this document are reserved No reproduction of this document is permitted w...

Страница 3: ...f the interlock system Test that all inputs to the L OEM handset are off and that buttons are working If the handset is awake hold buttons for 3 seconds The handset should go to the lower power state the display and all the LEDs should turn off To test the display backlights and left pushbutton wake the handset up by pressing the left pushbutton The display should show Alcolock Disabled OK to star...

Страница 4: ...lace ECU Communication error Handset has lost communication with ECU 1 Check voltage of vehicle during ignition cranking Voltage should stay above 9V 2 Replace spiral cable 3 Replace ECU 4 Replace Handset Error code 30 Handset blow tube temperature is out of range 1 Replace handset Error code 50 Sampling pump piston within the fuel cell is stuck in the active position 1 Replace handset Handset vol...

Страница 5: ...rd cylinder can be properly stored Transportation and storage of compressed gases in vehicles is dangerous and should be avoided Other precautions Never remove or alter canister labels Remove the valve and install the protective cap on cylinders when not in use Store cylinders in a cool well ventilated area away from sources of heat CYLINDER SETUP 1 Examine the canister and valve for any damage an...

Страница 6: ...reases in increments of 200 meters Press the right button to Accept 9 A Wait message is displayed followed by a 2 00 minute warm up The handset is ready for calibration when the TURN ON GAS message is displayed 10 Insert the delivery tube from the regulator valve into the mouthpiece port Keep holding the handset as the delivery tube is rigid 11 Press the valve down to release the gas while keeping...

Страница 7: ...The alcohol gas standard is expired A connection to or from the gas valve is not secured properly check all connections The value entered in Altitude menu is incorrect ALTITUDE REFERENCE GUIDE The altitude adjustment is in increments of 200 meters Choose the closest city elevation and round the altitude up or down to the closest value accordingly EXAMPLE CITY ALTITUDE M SETTING Amsterdam 0 0 Berli...

Страница 8: ...SG ENG 60 000566 A 2013 ...

Отзывы: