CALIBRATION MANUAL

3

NOTE: Press and hold the left button anytime to cancel time set and to return to the 

Date / Time screen.   

d. With the time adjusted, keep pressing Next until Set is displayed, and press Set.   
e. Press OK. The SELECT TYPE menu is displayed. 

8. 

Choose Dry Gas and press OK. The SELECT UNITS menu is displayed.

8. 

Select BrAC and press OK. The SELECT DIGITS menu is displayed.

9. 

Select 2 digits and press OK. The SELECT VALUE menu is displayed.

Read the next section. 

CALIBRATING A HANDSET (CONTINUED)

1. 

The SELECT VALUE menu contains the gas standard concentration options. Verify your 

canister label, select the appropriate PPM value, and press OK. This enters the Altitude 

menu.

Altitude is preset to 0 (meters).

2. 

Determine the altitude of your exact location.

NOTE:
•   Find exact altitude online 

•  Altitude may vary within general region or city; always use exact value 

3. 

If your altitude is: 

•  100 m or lower – leave at 0 m. Press OK
•  101 m or higher – press [+] to adjust value (increases by increments of 200 m). Round 

to the nearest increment (see table at back). Press Accept.

4. 

Wait message is displayed followed by a 2 minute warm-up. The handset is ready for 

calibration when the TURN ON GAS message is displayed.  

5. 

Insert the delivery tube from the regulator valve into the mouthpiece port. Keep hold of the 

handset as the delivery tube is rigid. 

6. 

Press the valve down to release the gas while keeping hold of the handset (see figure). 

Blowing message is displayed and the handset emits a tone.

Hold handset

Press valve

Содержание L OEM

Страница 1: ...Calibration manual Alcohol gas standard ...

Страница 2: ...s International Inc and are used under license Alcohol Countermeasure Systems is the trading style of Alcohol Countermeasure Systems Corp 2013 Alcohol Countermeasure Systems The information disclosed in this document is the valuable property of Alcohol Countermeasure Systems and all copyright and other proprietary rights to this document are reserved No reproduction of this document is permitted w...

Страница 3: ...d be avoided Examine the canister and valve for any damage and the expiration date on the label Observe all cautions and safety warnings found on the canister Never remove or alter canister labels Never modify the delivery tube in any way Always remove the valve and install the protective cap on cylinders when not in use Store cylinders in a cool well ventilated area away from sources of heat EQUI...

Страница 4: ...handset continued B BY MAINTENANCE CODE NO DOWNLOAD STATION NOTE The handset must be connected to the ECU 1 Power the handset on by holding down any navigation button 2 Access the menu screen by pressing and holding the left button NOTE Where available press and hold the right navigation button to exit a current screen menu 3 Press Next to scroll to System Maintenance and press Select 4 Input the ...

Страница 5: ...is enters the Altitude menu Altitude is preset to 0 meters 2 Determine the altitude of your exact location NOTE Find exact altitude online Altitude may vary within general region or city always use exact value 3 If your altitude is 100 m or lower leave at 0 m Press OK 101 m or higher press to adjust value increases by increments of 200 m Round to the nearest increment see table at back Press Accep...

Страница 6: ...calibration is complete If 3 or 4 tests are required a Calibration OK message is displayed when calibration is complete 8 Do I or II I For calibration with download station disconnect the handset from the download station and connect back to the ECU II For calibration without download station press the right button to finish there is also the option to retry by pressing the left button CALIBRATION...

Страница 7: ...rrectly entered in the Select Value menu The alcohol gas standard is expired All connections to and from the gas valve are not secured properly The value entered in Altitude menu is incorrect ALTITUDE REFERENCE TABLE The altitude adjustment is in increments of 200 meters Round the altitude up or down to the closest value as follows IF ALTITUDE IS THEN SET TO 0 100 m 0 101 300 m 200 301 500 m 400 5...

Страница 8: ...CAM ENG 60 000567 A 2013 ...

Отзывы: