REF: Alcolizer HH4 Alcohol Tester User Manual V1
Page 15
© Alcolizer Technology
Figure 16 - Ready for Passive Test Message
•
Press the Test button. The device will perform an electronic auto check and then
return to
Ready for passive Test
.
•
Hold the device 50mm in front of the test subjects mouth: so that the subject is
breathing or talking directly into the device’s sample inlet port (Refer to
Figure 17).
Figure 17 - Providing a Passive Sample
•
Ask the Subject to start counting from 20 onwards with a one second rhythm, until
you tell them to stop, “20…21…22…23…24…25…e
tc
.”
•
Press the Test button while the subject is counting. The device will make a series of
clicking sounds as the sample is being taken,
•
The device is set to take three pump samples. During these three pumps the
operator must hold the device steady and at the same distance from the subject.
The device will give three sample pump clicks, followed by a long beep sound.
Analysing Sample will then be displayed (Refer to Figure 18).