AlcoHAWK
®
Beacon Quick Guide
Be sure to read the entire owner’s manual and follow procedures
carefully before using this device.
Wait 20 minutes after the subject smokes, eats or drinks before testing.
Failure to wait 20 minutes can give inaccurate BAC readings and harm
the sensor.
Operating Instructions
Step 1)
Place a fitted cover over the mouthpiece each time you test another individual.
Step 2)
If it has been more than 24 hours since the device was last used,
prime the sensor
by blowing into the mouthpiece for 4 to 5 seconds
before
turning on the device.
Step 3)
PRESS
the Power Button and
HOLD
for
one (1) second
. A single beep will sound
and the device will display the room temperature.
Note:
The device will shut off after 30 seconds of non-use.
Step 4)
PRESS
the Left Button and the device will count down from 99 as it prepares the
sensor for testing.
Note:
The device will restart a count down from 10 if the sensor is not ready within the first countdown.
Step 5)
WAIT
until the
unit displays “rdY”
and you hear a
SINGLE BEEP
. Take a
DEEP
BREATH AND BLOW STEADILY
for at least 5 seconds. The unit will display “bL”
and count from 1 to 5 while you blow. A
SINGLE BEEP
sounds to signal completion
of the testing. Your estimated BAC will be displayed in the LCD and the backlight
indicates the alert level.
Note:
The device will shut off after 10 seconds of non-use.
DO NOT DRINK ALCOHOL AND DRIVE
SEE OWNER’S MANUAL FOR HELP UNDERSTANDING THE RESULTS, PRECAUTIONS AND RE-CALIBRATION INSTRUCTIONS FOR
RE-CALIBRATION THAT MUST BE ROUTINELY PERFORMED.
THIS DEVICE IS INTENDED TO MEASURE ALCOHOL IN THE HUMAN BREATH.
v03172014
FOLDING
MOUTHPIECE
BREATH OUT
Do not cover
while testing
LCD DISPLAY
With backlight
LEFT BUTTON
Select / Begin test
POWER BUTTON
OK / Confirm
RIGHT BUTTON
Recall last record /
Access user menu
BATTERY
COMPARTMENT
MOUTHPIECE HOLDER
Holds 2 mouthpieces