Alcofind AF-50 Скачать руководство пользователя страница 5

Replace new batteries. 

Replace new batteries 
immediately. 

Press           button again to 
repeat the measurement. 

Press           button to restart 
the countdown cycle and retest.   

Repeat the test after the unit 
has reached a temperature 
within the operation range.

Display message & Problem-solving guide

MESSAGE   

WHAT IT MEANS    

ACTION

Full battery

Empty battery  :  Warns when 
battery is low.   The unit can be 
operated but recommended to change 
new batteries for correct result!

If the LCD displays "BAt", 
the unit cannot operate.

If the sampling volume is too
low or not sufficiently blow.

If there is no breath sample blown 
within 30 seconds, the LCD will 
display "OUT".
If the unit is powered on beyond the 
acceptable temperature range(

) the unit will display it. 

5~40 C

or 41~104 F

Full battery

Technical specifications

Indication of B.A.C.

Accuracy

Warm up time

Sensor
Power supply
Battery Life
Dimensions(WxHxD)
Weight
Calibration
Operating temperature

0.000 ~ 0.500%BAC
0.00 ~ 5.00
0.00 ~ 2.50mg/l

+/- 0.005%BAC at 0.1%BAC at 25 C
+/- 0.05     at 1.0     at 25 C
+/- 0.025mg/l at 0.50mg/l at 25 C

Within 15 seconds at 0.50g/l(     )
warm up time may vary depending on the BAC measured.

Professional Fuel cell sensor
Two AA size alkaline batteries
Approx. 1500 tests
55mm x 125mm x 20mm
123g including batteries
every 12 months or after 500 tests
5 ~ 40 C

The specifications are subject to change without prior notice for functional improvement.

Содержание AF-50

Страница 1: ... cell sensor Personal professional use alcohol tester Instruction Manual 50 sets of data memory Professional Fuel Cell Sensor Law enforcement grade accuracy 4 digit LCD display with a backlit Fast analyzing and result Automatic power off User adjustable settings Blow time Unit Beep alarm Blow detecting ...

Страница 2: ...ve The manufacturer and the sellers will not take any legal responsibility for the test results When not used for a long time keep the device after taking out the batteries In case of testing after long term storage there might be high tolerance of test results for the first 3 tests Replace the batteries when the battery indicator icon reaches empty for the accuracy reason In case the test result ...

Страница 3: ...ower button The total number of test performed will be displayed briefly and start countdown When the countdown reaches zero with a double beep sound three dashes will be displayed and START will flash Take a deep breath and blow the mouthpiece for five seconds until there is a double beep sound Digital Breathalyzer ALCOFIND AF 50 Two AAsize Batteries 6 six Mouthpieces Carrying Case User s Manual ...

Страница 4: ...arm at 0 05 BAC the tester will sound an audible warning alarm when the test result is over 0 05 BAC If user set none the tester will not sound at all Recommended to set the value at your country s BAC Limit Adjust blow time When bL t appears on the display press the start button to adjust the blow time A longer blow time may provide a deep lung air which may provide more accurate reading Adjust f...

Страница 5: ...lown within 30 seconds the LCD will display OUT If the unit is powered on beyond the acceptable temperature range the unit will display it 5 40 C or 41 104 F Full battery Technical specifications Indication of B A C Accuracy Warm up time Sensor Power supply Battery Life Dimensions WxHxD Weight Calibration Operating temperature 0 000 0 500 BAC 0 00 5 00 0 00 2 50mg l 0 005 BAC at 0 1 BAC at 25 C 0 ...

Страница 6: ...bration under normal use either for 1 one year from the date of purchase or 500 breath tests whichever comes earlier The manufacturer s obligation under this warranty is limited to replacing adjusting or repairing the unit at manufacturer s option if returned along with proof of purchase For any defects due to natural disaster or customer s negligence misuse reconstruction and repair by unauthoriz...

Отзывы: