background image

If you do not wait 20 minutes before testing, the reading will measure the alcohol in your mouth, 
which is many times that related to your blood alcohol level. The detector contains a highly sensitive 
semi-conductor sensor, and not waiting the required time could damage (saturate) this sensor and 
can cause all future readings to be inaccurate. Please also ensure you do not blow any smoke or 
liquid into the unit, or the same damage can result and de-calibrate the detector. 

 

Press the ON/OFF power switch.  The detector will start to count down from 200 to 000 on the 
display.  The warm up and cleaning cycle will prepare the sensor for testing. 
 
After hearing a BEEP and the green READY light comes on, take a deep breath, then gently and 
steadily blow into the breath intake until you hear the second BEEP.  You will blow for 3 seconds.  
Make sure your lips are sealed around the pipe and your hand has not covered the air outlet. 
 
After the red and green display lights blink for 4 seconds, the test results will be displayed by 3 digits 
and remains for 15 seconds. 
 

Time out: 

if you don’t blow within 30 sec from READY, the unit will time out and displays a blinking 

“OFF”. Press the ON/OFF power switch to turn the unit off and repeat step 1. 

 
How to Analyze the Results?

 

   If the reading is over 0.05, the red WARN lamp will flash along with a BEEP sound. 

 

   If the sensor is saturated ,the display reads “HOT” and the red lamp will   flash along with 

the BEEP sound. This can occur if a subject blows alcohol from their saliva into the tester. 
 

   If you consume very little alcohol, your actual concentration may be below .01% BAC and 

the reading may not be activated. However, the display will read 0.00 
 

   The UK Drink-Drive limit is 0.08% BAC – displayed as 0.08 on the CA2000 

 

 

Precautions

 

1. 

Always wait at least 20 minutes after your last drink before testing

. This is because it 

can take up to 20 minutes for any residual alcohol in the mouth to dissipate and for alcohol to 
be fully absorbed in the blood. 

2. 

Avoid testing in strong winds, in a closed room with a heavy smoke, or where a lot of alcohol 
is being consumed. 

3. 

Do not blow cigarette smoke, food or liquid into the instrument  

4. 

The  CA2000 PRO is designed for use in a temperature range of 10 – 40 (C) or 50-104 (F). 
Operation of the unit in extreme temperature ranges above or below this range may affect 
the accuracy of results. 

5. 

Avoid testing in the presence of any substances that contain methyl alcohol, isopropyl 
alcohol or acetone. These substances may interfere with the results of the test. 

6. 

Conditions that increase the amount of ketenes on the breath, such as diabetes and low 
caloric intake, may cause a false positive test. 

7. 

DO NOT DRINK AND DRIVE. Always have a designated driver if you are drinking 
Alcohol. 

Always leave at least 20 minutes between drinking and testing 

How does it work? 

When you drink alcohol it passes through the lining of the stomach into the blood supply.  As the 
blood moves around your body it is filtered by the liver (which extracts the alcohol over a period of 
time) but also passes through the lungs. At the same time as the oxygen in your breath passes 
through the alveoli into your blood, a certain amount of the alcohol that is carried by the blood passes 
back the other way. It is therefore important to measure deep lung air as this is where the blood has 
been in contact with the air in the lungs. A breath from any other part of the lungs will have a lower 
alcohol concentration and can give a lower reading than should be the case. 
 
To achieve a consistent reading, a consistent sample of deep lung air is therefore required.  Law-
Enforcement devices measure an absolute quantity of air (usually 1.7ltr) each time they are used and 
employ a chemical fuel cell to measure alcohol concentration. This however makes such devices 
significantly more expensive to manufacture than personal detectors like this that use a “timed blow” 
& semi-conductor sensor, but with practice the user can achieve a reasonable level of accuracy by 
ensuring that as far as possible the same blow pressure is used each time a sample is taken.   

 
 
How to get the most from your Breathalyser

 

When first using the device, 20 minutes after your last drink try testing yourself three times, 
approximately two minutes apart, in order to “bracket” the readings. An average reading can then be 
assessed, as can the consistency of the blow level required.   Note the reading, and then test again 
30 minutes later and so on until the unit reads zero.   
 
Try the unit on others; you are likely to find considerable variance in the way in which individuals 
metabolise alcohol, depending upon body mass, when they last ate, time of day etc. After a while you 
will start to develop a reasonable idea of the way in which you personally react to Alcohol and you 
can then modify your drinking appropriately to ensure that you always leave a more than adequate 
period after drinking and before driving. 
 
UKBreathalysers would always recommend you do not drive with ANY alcohol in your system, and 
that you leave a substantial safety margin between the time you believe you have reached a zero-
level and the point where you may consider driving. 

Never rely solely upon the readings of this 

detector  in determining your fitness to drive! 

 
 
Disposable Mouthpieces 

Your detector is equipped with removable / disposable mouthpiece. For all sanitary considerations 
replace and dispose the mouthpiece when testing others. To reorder please visit our online store or 
call our order desk.  

 
 
Calibration Procedure 

If your device is providing inconsistent, unusually high or low readings, or no readings at all, your 
device may need to be re-calibrated. In addition, the breathalyser

 

should be re-calibrated at least 

once every 6-12 months. More frequent calibrations may be required depending on frequency of use. 

  

For detailed information on how to have the device re-calibrated, please see our Website. 

 
 
Cleaning 

Although it may seem logical to clean the alcohol detector with alcohol or other cleaning agents, in 
reality, this is very harmful to the semiconductor sensor in detector. All semiconductor sensors are 
intolerant to high doses (saturation) of alcohol, and other cleaning or anti bacterial agents.

   

Simply 

wipe the unit over with a damp (not wet) cloth and dry thoroughly. 

CA2000 

User Manual 

Отзывы: