![Albrecht CDS MANUSAN Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/albrecht/cds-manusan/cds-manusan_user-instructions_2889660010.webp)
10
CDS
®
Manusan
User Instructions
2.1.3. Adjusting the position of the redression and limitation range
The
CDS
®
Manusan
is constructed on the basis of a 360° hinge. In a 360° hinge,
the position of the spring housing is adjustable in relation to the forearm hinge
rod. The position of the redression and limitation range can be thereby altered
and adjusted to the patient. The position of the stops depends on the position
of the redression and limitation range.
1
Remove the pad.
2
Remove the two screws.
3
You can adjust the position of the redression range in 15° steps. Use
a goniometer to determine the desired position and adjust the hinge
accordingly.
4
Insert the screws again and tighten them.
5
Insert the pad again.
That way, you can adjust the range of motion of the brace to the patient even if
there are extreme motion deficits.
1
2
3
4
5