When users adhere to 

Alaxo’s

 recommendations, the cleaning procedures provide proper stent 

maintenance as well as cleaning and disinfection of installation equipment.

 

After inserting the stent(s) into the nose and releasing the stent(s) from the pusher rod, rinse the 
insertion tube and pusher rod in 

COLD

 water until all attached mucus is removed. Spray the outer 

surface of the insertion tube with cleaning/disinfection spray (Protosan® spray) and spray into the 
end of the insertion tube. This allows the spray to coat the inside of the tube. Allow the spray to 
remain for one minute, rinse with lukewarm water and let the insertion tube air dry.

 

After removing the stent(s) from the nose, rinse the stent(s) in lukewarm water to remove attached 
mucus.

 

Use lukewarm running water and apply your thumb and forefinger in a stroking motion, from the 
silver ball end to the end of the stent, to gently clean remaining mucus. Alternatively, a soft bristle 
brush, such as a toothbrush, may be used. Caution: There is a potential for damaging the stent if the 
thumb and fingers or brush is used against the grain of the stent braid.

 

When the stent(s) are clean, spray them with the cleaning/disinfection spray (Protosan® spray) and 
allow the spray to remain for one minute. Then rinse with lukewarm water and allow stent(s) to air 
dry.

 

To determine if the stent(s) are contaminant free, hold the stent up to a bright light and look closely at 
the braided wires to see if any lint or mucus residue remains. If contaminants exist, repeat cleaning 
process. The proper removal of all contaminants from the stent(s) prevents potential irritants affecting 
the nasal tissues.

 

Storing your stents

 

Do not store the stent compressed in the insertion tube as this creates unnecessary load pressure on 
the stent that diminishes 

the stent’s

 ability to reopen to its proper shape.

 

The recommended storage temperatures range should be above 4o C (39o F) and less than 50o C 
(122o F) and should not be stored in direct sunlight.

 

Protosan® information

 

The cleaning/disinfection spray consisting of 0.1% polyhexanide / 0.1% betaine reaches

 

all the small surface areas because of the spray’s modified surface tension. The cleaning/disinfecting

 

spray ingredients possess excellent mucosa (nasal tissue) compatibility and provide a broad-spectrum 
pathogen disinfectant.

 