Automatic reaction of ADS intelligent
overpressure (+30%) : the system will automatically control valves to decrease the pressure
level inside the mattress if necessary.
Alternating/cycle time mode: this mode will automatically be enabled after passing by the
first interval of patient monitoring a) without redistributing any movement of patient (please
see chapter
Error! Reference source not found.
for details) AND b) "Alternating/cycle time
mode" is enabled.
One cycle of alternating will be proceed as follows:
1) deflating the left or right row of small cells for approximately 10 minutes.
2) deflating the middle row and inflating the previous empty row for approximately 10
minutes.
3) inflating the middle row and deflating the last remaining row right or left (in opposite to the
row in 1)) for approximately 10 minutes.