5
3. Fill the bottom section of the nebulizer
with the medication prescribed by your
Physician. Be sure that the cone is put
Place cone inside
inside the bottom section.
and fill medication
4. Gently twist the top part clockwise to
reassemble the nebulizer. Be sure the
two sections fit well.
5. Attach one end of air tube to the base of the nebulizer.
6. Attach the other end of air tube to the air tube connector located on the back
of the compressor.
7. Attach the interface accessory of your choice to the top section of nebulizer.
8. Plug the power cord into an appropriate electrical outlet. Make sure at this
stage, the power switch is at “OFF” status.
9. Press the power switch to “ON” status to begin your prescribed treatment.
Important:
The compressor motor has a thermal protector, which will shut off the unit
before the unit is overheated. When the thermal protector shuts the unit off,
please:
a. Switch off the unit.
b. Unplug the unit from the electrical outlet.
c. Wait 30 minutes for the motor to cool down before another treatment.
d. Make sure the air openings are not obstructed.
10. When treatment is finished, shut off the unit and unplug it from the electrical
outlet.
5.
CLEANING
It is recommended that the nebulizer and the mask (or any other patient
accessory interface that is used) be thoroughly cleaned with hot water after each
use and cleaned with a mild detergent after the last treatment of the day. If your
physician or respiratory therapist specifies a different cleaning procedure, follow
their instructions.
Caution:
Before cleaning, make sure the power cord is disconnected from the main outlet
and the unit.
Rinsing (after each treatment)
1. Disconnect the air tube from the nebulizer. Disconnect the mask or other
patient accessory interface from the nebulizer.
2. Gently twist the top part of the nebulizer counter-clockwise to disassemble