Airial MQ3000 Скачать руководство пользователя страница 2

Declaration 

Guidance and Manufacturer

’s declaration – electromagnetic emissions- 

For all EQUIPMENT and SYSTEMS 

Guidance and Manufacturer

’s declaration - electromagnetic emission 

The MQ3000 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user 
of MQ3000 Pulse Oximeter should assure that it is used in such an environment. 

Emission test 

Compliance 

Electromagnetic Environment 

– guidance 

RF emissions CISPR 11 

Group 1 

The  MQ3000  Pulse  Oximeter  uses  RF  energy  only  for  its 
internal  function.  Therefore,  its  RF  emissions  are  very  low 
and  are  not  likely  to  cause  any  interference  in  nearby 
electronic equipment. 

RF emissions CISPR 11 

Class B 

The  pulse  Oximeter  (MQ3000)  is  suitable  for  use  in  all 
establishment, including domestic establishments and those 
directly  connected  to  the  public  low-voltage  power  supply 
network that supplies buildings used for domestic purposes. 

Harmonic emissions IEC 61000-3-2 

Not Applicable 

Voltage fluctuations/ flicker emissions 
IEC 61000-3-3 

Not Applicable 

Guidance and Manufacturer

’s declaration – electromagnetic immunity- 

For all EQUIPMENT and SYSTEMS 

Guidance and Manufacturer

’s declaration - electromagnetic immunity 

The MQ3000 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user 
of the MQ3000 Pulse Oximeter should assure that it is used in such an environment. 

Immunity test 

IEC  60601  test 
level 

Compliance 
Level 

Electromagnetic Environment 

– guidance 

Electrostatic 
Discharge (ESD) 
IEC 61000-4-2 

+/- 6kV contact 
+/- 8kV air 

+/- 6kV contact 
+/- 8kV air 

Floors  should  be  wood,  concrete  or  ceramic  tile.  If 
floor are covered with synthetic material, the relative 
humidity should be at least 30%. 

Power frequency (50/60 Hz) 
magnetic field 
IEC 61000-4-8 

3A/m 

3A/m 

Power frequency magnetic fields should be at levels 
characteristics  of  a  typical  location  in  a  typical 
commercial environment. 

Guidance and Manufacturer

’s declaration – electromagnetic immunity- 

For all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING 

Guidance and Manufacturer

’s declaration - electromagnetic immunity 

The MQ3000 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user 
of the MQ3000    Pulse Oximeter should assure that it is used in such an environment. 

Immunity 

test 

IEC  60601  test 

level 

Compliance 

Level 

Electromagnetic Environment 

– guidance 

Radiated 

RF 

IEC 

61000-4-3 

3 V/m 

80  MHz  to  2.5 

GHz 

3 V/m 

Portable  and  mobile  RF  communications  equipment  should  be 
used  no  closer  to  any  part  of  the  Pulse  Oximeter  (MQ3000), 
including  cables,  than  the  recommended  separation  distance 
calculated  from  the  equation  applicable  to  the  frequency  of  the 
transmitter. 
Recommended separation distance 

  80 MHz to 800 MHz 

  800 MHz to 2.5 GHz 

Where P is the maximum output power rating of the transmitter in 
watts (W) according to the transmitter manufacturer and d is the 
recommended separation distance in meters (m). 
Field  strengths  from  fixed  RF  transmitters,  as  determined  by  an 
electromagnetic site survey

a

, should be less than the compliance 

level in each frequency range. 

b

 

Interference  may  occur  in  the  vicinity  of  equipment  marked  with 
following symbol: 

 

NOTE 1    At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2    These guidelines may not apply in all situations, Electromagnetic propagation is affected by absorption and reflection 
structures, objects and people. 

a    Field strengths from fixed transmitters, such as base station for radio (cellular/cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic  environment  due  to  fixed  RF  transmitters,  an  electromagnetic  site  survey  should  be  considered.  If  the 
measured field strength in the location in which the Pulse Oximeter (MQ3000) should be observed to verify normal operation. If 
abnormal performance is observed, additional measurements may be necessary, such as reorienting of the relocating the Pulse 
Oximeter (MQ3000). 
b    Over the frequency range 150 kHz to 80 MHz, fields strengths should be less than 3 V/m 

Recommended separation distances between portable and mobile RF communications equipment and   

the EQUIPMENT or SYSTEMS - For all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING 

Recommended separation distances between   

portable and mobile RF communications equipment and

 Pulse Oximeter (MQ3000)

 

The 

Pulse  Oximeter  (MQ3000)

  is  intended  for  use  in  electromagnetic  environment  in  which  radiated  RF  disturbances  are 

controlled.  The  customer  or  the  user  of  the 

Pulse  Oximeter  (MQ3000)

  can  help  prevent  electromagnetic  interference  by 

maintaining  a  minimum  distance  between  portable  and  mobile  RF  communications  equipment  (transmitters)  and  the 

Pulse 

Oximeter (MQ3000) 

as recommended below, according to the maximum output power of the communications equipment.   

Rated  maximum  output 

power of transmitter (W) 

Separation distance according to frequency of transmitter (m) 

80 MHz to 800 MHz 

 

800 MHz to 2.5 GHz 

 

0.01 

0.1167 

0.2334 

0.1 

0.3689 

0.7378 

1.1667 

2.3334 

10 

3.6893 

7.3786 

100 

11.6667 

23.3334 

For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can 
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of 
the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people. 

Possible Problems and Solutions 

Problems 

Possible reason 

Solution 

SpO

2

 or PR can not 

be shown normally 

1. Finger is not inserted correctly 
2.The user

’s Oxyhemoglobin value is too low to be 

measured 

1.Retry by inserting the finger 
2. Try not to move 

 

 

 

SpO

2

 or PR is shown 

unstably 

1. Finger might not be inserted deep enough. 
2. Finger is trembling or user

’s body is in movement 

status. 

1. Retry by inserting the finger 
2. Try not to move 

The oximeter can not 
be powered on 
 

1. Power of batteries might be inadequate or not be 

there at all. 

2. Batteries might be installed incorrectly. 
3. The oximeter might be damaged. 

1. Please replace batteries 
2. Please reinstall the batteries 
3. Please contact with local customer service 
centre 

Indication lamps are 
suddenly off 

1. The product is automatically powered off when no 

signal is detected longer than 8 seconds 

2. Power quantity of the batteries is started being 

inadequate 

1. Normal   
2. Replace the batteries 

“Error3”    or   

“Error4” is displayed 
on screen 

1. Low power   
2. Receiving tube being shielded or damaged together 

with broken connector. 

3. Mechanical Misplace for receive-emission tube. 
4. Amp circuit malfunctions. 

1. Change batteries 
2. Please contact local customer service center 
3. Please contact local customer service center 
4. Please contact local customer service center 

Error 6 

Err 6 means the screen is failure 

Change the screen 

“Error7” is displayed 
on screen 

1. Low power   
2. Emission tube damaged.   
3. Current control circuit malfunctions. 

1    Please change battery   
2    Please contact local customer service center 
3    Please contact local customer service center 

Symbol Definitions 

Symbol 

Definition   

 

Type BF applied part. 

 

Attention, consult accompanying documents. 

 

Protected against dripping water. 

SpO

2

 

Oxygen saturation   

 

Pulse rate (BPM) 

 

Low power indication   

 

No SpO

2

 Alarm 

 

Power switch 

SN 

Serial No. 

 

Storage temperature and relative humidity 

 

Manufacturer’s information 

 

Date of Manufacture 

 

European union approval 

 

Authorized representative in the European community 

Note: The illustrations used in this manual may differ slightly from the appearance of the actual produc 

 

Manufactured For: 

 

 

 

P.O. Box 1559 
Bluffton, SC 29910 
(888). 404.5666 

 

 

 

 

ALL RIGHTS RESERVED 

Issue Date: 2/March 2012

 