![Airflow Instruments PAN300 Series Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/airflow-instruments/pan300-series/pan300-series_operation-and-service-manual_2876831037.webp)
31
Appendix D
Procedure for Using Smoke Pellets in
Leakage Testers
As shown in the sketch below, a rubber bung is fitted into the hole in
the duct adapter which holds the wire coil.
When a smoke pellet is required to be used, remove the bung and fit
a pellet into the wire coil as shown in the sketch.
Light the pellet and immediately plug into the hole in the duct adapter
and proceed with the test.
The pellet should emit dense white smoke for up to a minute.
Duct
Pellet Holder
Smoke Pellet
Flexible Ducting
Duct Adapter