1. Setting up and using the Variflow pump

Before you use the VARIFLOW pump make sure it is fully charged. A battery level indicator 
shows the available power when the unit is first switched on.

1.1  Charging the battery

When the charger is plugged into the charging socket a red LED will be illuminated. The LED 
will change to green when the battery is fully charged. It takes approximately 6-8 hours to 
completely charge the battery.

1.2  Removing the waterproof covers

IMPORTANT: Before switching the unit on, ensure the gold air outlet valve and sampling 
head are not covered.

2. Switching on and setting the flow rate 

Before switching on the VARIFLOW connect a flow meter to the air sampling head.
Press 

start/pause

 to switch on the unit. The display will show, in this order, the serial 

number, the current firmware version and the battery status.
If 

start/pause

 is pressed the display shows ‘Flow not set’.

Press either arrow to open and start Calibration Mode and to Set Flow. 
The pump is set at aprox 8 l/min as a default, to fine tune this:-
Press 

start/pause

 to start calibration. NOTE: Run for 45 seconds to allow to stabilise.

Press up/down arrows to adjust flow rate. (A single press advances the display by one 
increment, holding while pressing advances the display by 10 increments). Check the 
flow rate against the flow meter and adjust acording to flow requirements. (Any increment 
changes will be saved for the post test flow check, see 2.2 & 2.3).
To confirm and set the flow rate press 

stop

.

IMPORTANT: If pump is shutdown between setting the flow and starting the test, flow will 
have to be re-calibrated by repeating the steps above.

2.3 If Timed Test is selected 

Press 

start/pause

 the display shows ‘Enter Run Time’

Press arrows to adjust time:- 
Press 

start/pause

 to move between (Hrs) 00 (Mins) 00 (Sec) 00.

Press 

start/pause

 the pump runs and the timer starts. When the time has elapsed the 

display shows ‘Test Completed - Check Flow’.
Press 

stop

 display shows ‘Test completed - Check Flow’.

Press 

stop

 display shows ‘Connect flow meter’ (now reconnect your flow meter to check 

the flow rate).
Press 

start/pause

 display shows ‘Flow Check’ (any increment changes made when 

setting the flow rate (see 2) will be shown. 

NOTE: Wait 45 seconds for flow rate to stabilise.

Press 

stop

 display shows ‘Flow Check Complete’

Press 

stop

 to shutdown.

Note:  If the pump switches itself off during either Free Running or Timed Test (i.e if the 

battery dies or the sampling head gets blocked) on restart it will either show:-

1. ‘Test Incomplete, Restart Test?’, 

Press 

start/pause

 to resume test and see where the 

previous test got to.

2. ‘Test Completed Check Flow’, in which case continue as instructions above.
3. ‘Low battery’, in which case either replace battery (within 10 minutes of disconnecting 

old one) or charge pump directly from the mains for a short time to complete the flow 

check. 

2.2  If Free Running is selected 

Press 

start/pause

 the pump runs and the timer starts counting, (you can pause during 

the test by pressing 

start/pause

)

Press 

stop

 during test and the display shows ‘Test Stop? Are you sure?’ If you are not sure, 

press 

start/pause

 to continue test, if you are sure:-

Press 

stop

 display shows ‘Test completed - Check Flow’.

Press 

stop

 display shows ‘Connect flow meter’ (now reconnect your flow meter to check 

the flow rate).
Press 

start/pause

, display shows ‘Flow Check’ (any increment changes made when 

setting the flow rate (see 2) will be shown). 

NOTE: Wait 45 seconds for flow rate to stabilise.

Press 

stop

 display shows ‘Flow Check Complete’.

Press 

stop

 to shutdown.

2.1  Setting test mode

Press 

start/pause

 the display will show ‘Set test mode’.

Press arrows to select ‘Free Running’ or ‘Timed test’.