Trouble

Possible Cause

Corrective Action

1 No power No green 

light on membrane 
switch

• Not plugged in.

• No power at Mains Outlet

• ScanX has not been turned on.

• Defective power supply

• Check the line cord connection is 

firmly plugged in .

• 

Make sure outlet is grounded and has 
power .

• Make sure that the main power 

switch is set to ON (1) .

• Call your Air Techniques dealer.

2 Green, Blue or Yellow 

indicator does not 
light .

• Defective light or circuitry.

• Call your Air Techniques dealer.

3 Image Management 

Software does not rec-
ognize ScanX when 
selected .

• Inadequate Computer System.

• ScanX has not been turned on.

• The computer connection cable 

is loose or defective .

• 

The computer does not recognize 
that ScanX is connected .

• ScanX hardware problem. 

• Verify  Computer  requirements 

(Page 9) .

• 

Make sure that the POWER switch is 
set to ON and the green indicator 
light is lit .

• Reconnect  the  cable.  Check  for 

tightness . Replace if necessary .

• 

Verify that the Setup program was 
correctly installed (Page 13) .  Use 
different USB port .

• Call your Air Techniques dealer.

4 Plate does not scan 

properly .

• The PSP was not pushed far 

enough into ScanX .

• Fully  feed  the  PSP  into  the  Plate 

Guide .

5 No image appears 

after scanning .

• The PSP fed backwards

 

(printed side towards ScanX) .

• The PSP was erased prior to 

scanning .

• X-ray source failed.
• Hardware failure.

• 

Quickly refeed the plate with the 
printed side out .

  If Eraser mode 

was enabled 

during scanning, you may need 
to retake the image .

• 

Feed the PSPs into the scanner 
immediately and quickly after 
removal from the Barrier Envelope .

• Call your X-ray service dealer.
• Call your Air Techniques dealer.

Important: 

Do not allow the PSP to 

be exposed to light between 

taking an X-ray and scan-

ning with the ScanX.

Содержание ScanX Duo

Страница 1: ...ScanX Duo Digital Imaging System Instruction Manual...

Страница 2: ...s a leading manufacturer of dental medical and veterinary equipment since 1962 The ScanX Duo is designed to process intraoral imaging plates only See the sections Unpacking and Inspection and Technica...

Страница 3: ...or symbols are used on the equipment on the serial number label or in this manual to denote information of special importance DANGER AVOID DIRECT EXPOSURE TO BEAM LASER RADIATION WHEN OPEN CAUTION CLA...

Страница 4: ...of ScanX is only accessible by removing hardware with tools It should be opened and ser viced only by an authorized dealer service technician Failure to heed this warning may result in equip ment dam...

Страница 5: ...a trained technician from an authorized dealer should remove a cover from ScanX Direct eye contact with the output beam from the laser may cause serious damage and possible blindness Do not open ScanX...

Страница 6: ...ich accumulate condensation or become wet through a change of temperature may only be operated after they are fully dry Air Techniques regard themselves as being responsible for the equipment with reg...

Страница 7: ...e The manufacturer is not to be held liable for any damage caused as a result of incorrect usage The operator bears all risks This manual provides the information necessary for the setup operation and...

Страница 8: ...item is missing notify your dealer Unpacking and inspection System Components Description Quantity Part No 10 Foot Power Cord 1 73096 6 Foot 2 0 USB Cable 1 B3554 Accessory Kit containing Pack of 4 S...

Страница 9: ...rvice Pack 3 or later for an Intel 32 bit processor Microsoft Windows XP Professional 64 bit Edition with Service Pack 2 or later for an Intel 64 bit extended x64 processor Microsoft Windows Vista Bus...

Страница 10: ...indicator Illuminates steady blue to indicate that the Erase function is ON and PSPs will be erased after scanning Flashes blue to indicate that the Erase function is OFF and PSPs will not be erased...

Страница 11: ...d Resolution Range Horizontal 10 to 12 lp mm Vertical 10 to 18 lp mm Compliance Data Laser Class Class I Laser Product 21 CFR 1040 10 Class 1 Laser Product IEC 60825 1 Laser Product Report Accession N...

Страница 12: ...s AC or a c alternating current CD ROM compact disk read only memory CFR Code of Federal Regulations CPU central processing unit your computer cm centimeter D depth GB gigabyte 230 109 bytes GHz Giga...

Страница 13: ...e computer meets all requirements listed on page 9 3 Verify that an authorized Imaging Software is installed properly on the computer 4 Verify that the supplied ScanX Drivers and Utilities Disk which...

Страница 14: ...Press HERE to turn ON or OFF Standby Main Power Switch 1 0 See Note IEC Connector 2 A Circuit Breakers TO AC OUTLET Line Cord Keypad Control Panel USB Connector Type B USB TO COMPUTER POWER Switch ER...

Страница 15: ...must always be erased prior to use Note Use PSPs within 24 hours of last erasure Repeat erasing process if PSPs have been stored longer than 24 hours Erase the PSP Each Intraoral PSP should be used i...

Страница 16: ...Plate surface Wipe back and forth and then in a circular motion 2 Allow the plate surface to air dry Make sure that the PSP is completely dry before re using Disinfecting the Phosphor Storage Plates T...

Страница 17: ...lates in the X ray session have been exposed prepare each plate by performing the procedure below Preparing the Exposed Plate for Scanner Processing 1 Disinfect the Barrier Envelope with plate still i...

Страница 18: ...th the green and blue LED indicators above the POWER and ERASER switches respectively illuminate Default mode has eraser ON 4 Use the user supplied authorized Imaging Software to activate the Scanner...

Страница 19: ...when removing from the scanner outlet 10 View and save each scanned image using the user supplied authorized Imaging Software IMPORTANT PSPs will not be erased after scanning when operating ScanX wit...

Страница 20: ...fect the Plate Guide unless contamination is suspected If a Plate Guide has touched a contaminated surface it may be immersed briefly in a cold sterilant according to sterilant manufacturers direction...

Страница 21: ...or white side remains relatively smooth This scratched look on the black side has absolutely no effect on the quality of the image and should be expected under normal conditions If the phosphor side...

Страница 22: ...m Verify Computer requirements Page 9 Make sure that the POWER switch is set to ON and the green indicator light is lit Reconnect the cable Check for tightness Replace if necessary Verify that the Set...

Страница 23: ...ge due to exposure to light during handling of the PSP Make sure ScanX is operating with eraser turned on blue LED indicator above the ERASER switch is illuminated steadily Make sure the plates are in...

Страница 24: ...s Size Order No Qty 0 73248 0 100 1 73248 1 100 2 73248 2 300 3 73248 3 100 4 73248 4 50 ScanX Cleaning Sheets Description Order No Qty ScanX DUO Sheets D1010 12 ScanX DUO Sheets D1020 25 One re usabl...

Страница 25: ...al imaging ScanX Systems and Polaris Intraoral Cameras Key Features Easy to learn and simple to use Acquire mount view and store images with just a few mouse clicks Customize images to personal prefer...

Страница 26: ...replaced at our option at no charge provided that our inspection shall indicate it to have been defective and that the system is returned to our factory properly received Dealer labor shipping and han...

Страница 27: ...care professionals only This equipment may cause radio interference or may disrupt the operation of nearly equipment It may be necessary to take mitigation measures such as reorienting or relocating t...

Страница 28: ...ine s 2 kV line s to earth 1kV N A Mains power quality should be that of typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines...

Страница 29: ...rom fixed RF transmitters as determined by an electromagnetic site survey a should belessthanthecompliance level in each frequency b Interference can occur nearby devices bearing the following symbol...

Страница 30: ...er W Separation distance as a function of the transmitter s frequency m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 16 P d 1 16 P d 2 33 P 0 01 0 116 0 116 0 233 0 1 0 366 0 366 0 736 1...

Страница 31: ...Page 31 Air Techniques Inc Notes...

Страница 32: ...hnical support Whether the need is digital imaging utility room equipment or merchandise Air Techniques can provide the solution via our network of authorized professional dealers Proudly designed tes...

Отзывы: