background image

Mojave LT Motor Replacement Kit

Part No: H4150 and H4151 

Replacement 

Instructions

© 

2019 Air Techniques, Inc. • P/N H4150-LIT, Rev. B - March 2021

1295 Walt Whitman Road

 

|

 

Melville, New York 11747- 3062

Phone: 800-247-8324

 

|

 

Fax: 888-247-8481

Website: www.airtechniques.com

Introduction

This document provides instructions for using replacement kits (P/N H4150 and H4151)

 

to replace a motor assembly of a 

MOJAVE LT Dry Vacuum System.  

Important Information

Service Personnel

 

All personnel servicing the MOJAVE LT system must read and understand the Operator's Manual 

before operating or performing any service of the unit.  Only a trained technician from a qualified Air 

Techniques dealer should perform any service. 

Power Removal

 

Never work on equipment connected to live power.  Remove power by disconnecting mains 

power to the unit prior to performing any replacement procedures.

Motor Replacement / Refurbished Kit Components Supplied

1. 

Turn power to the device OFF.

2. 

Remove any hoses connected to the motor exhaust port.

3. 

Disconnect the brown MOTOR and orange MOTOR FAN circular connectors from the back of the electrical box by 

loosening the locking ring closest to the electrical box on each connector.

Motor Replacement Procedure

Replace the faulty motor using the replacement or refurbished motor by performing the following steps.  

Motor Assembly

A

Mojave LT Front View

Motor 

Exhaust Port

Motor Exhaust Connection

Mojave LT Rear View

MOTOR and MOTOR FAN Connections

Orange

Brown