14
2109100018L29 1608V002
Product description
5.1 Cam interchangeable head
1
2
3
1 Optical system
2 LED
3 LED
The Cam interchangeable head has an optical
element with autofocus with a focal range for in-
traoral recordings When placing on the inter-
changeable head, the focus level is preset to
two molars Positioned around the optical ele-
ment are two LEDs that ensure uniform illumina-
tion
Figure 1: Recording with Cam interchangeable
head
The interchangeable head is plugged onto the
handpiece and connected via the contacts A
guide prevents incorrect placement of the inter-
changeable head
On each side of the handpiece there are two
buttons: the focus button and the capture but-
ton The pressure point of the buttons is notice-
able
The focus button is used to focus the camera
sharply on the object If the Spectra inter-
changeable head is put on, the focal plane is
preset on the spacer but can be changed with
the focus button
Still images and video recordings can be creat-
ed with the camera The function of the trigger
button depends on the recording mode in the
imaging software (still image or video) In the Still
Image mode, the camera switches between Live
mode (moving image) and Freeze mode (still im-
age) In Video mode, the recording starts or
stops Releasing the trigger button causes the
camera to vibrate slightly Optionally, a foot
switch can also be used for triggering
The illumination is incorporated in the inter-
changeable head The optical element is divid-
ed: One part is in the handpiece, the other part
is in the interchangeable head
The image sensor in the handpiece digitizes the
image The camera transmits the image to a
computer via the USB connection cable
The connection cable is used to connect the
camera directly to the USB connection of the
computer or, optionally, to the handpiece holder
with USB hub To run the camera an imaging
software is necessary The Spectra interchange-
able head require imaging software from Air
Techniques or a TWAIN compatible imaging
software
The power supply of the camera to the comput-
er is realized via the USB connection cable
The camera switches off automatically if it is not
moved for two minutes As soon as the camera
is moved, it switches on again
EN
US